Report output
N O T I C E
The data output (printer or PC) needs to be selected first, so that the Type of
report menu is available (refer to
Select the data output
on page 11).
Different report output types can be selected when a printer or a PC is
connected.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Type of report and confirm.
3.
When a printer or a computer and Terminal is connected, use the
or to select
Option
Description
Reduced
Select Several or One sample as an output format
Standard
Select Several or One sample as an output format. Select
several:
Users:
The user name appears on the printed report
(17 characters).
Header:
The company name can be added as a
header (40 characters) and appears on the print report.
Identify
sensor:
The sensor model and the sensor serial number can be
added and appears on the print report.
GLP
Select Several or One sample as an output format. Select
several:
Users:
The user name appears on the printed report
(17 characters).
Header:
The company name can be added as a
header (40 characters) and appears on the print report.
Identify
sensor:
The sensor model and the sensor serial number can be
added and appears on the print report.
4.
When a computer is connected and LabCom Easy (refer to
Select
the data output
on page 11 for more information) is selected, use the
or to select
Option
Description
Users
The user name appears on the printed report
(17 characters).
Identify sensor
The sensor model and the sensor serial number can be
added and appears on the print report.
Maintenance
W A R N I N G
Multiple hazards. Do not disassemble the instrument for maintenance or service.
If the internal components must be cleaned or repaired, contact the
manufacturer.
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.
Clean the instrument
N O T I C E
Never use cleaning agents such as turpentine, acetone or similar products to
clean the instrument including the display and accessories.
Clean the exterior of the instrument with a moist cloth and a mild soap
solution.
Clean the probe
Clean the probe as needed. Refer to
Troubleshooting
on page 16 for
more information about cleaning. Refer to the probe documentation for
information about the probe maintenance.
Use the cleaning agents listed in
Table 1
for contaminations on the pH
probe.
Table 1 Cleaning agents for the pH probe
Contamination
Cleaning agent
Proteins
Pepsin cleaning solution
Grease, oils, fats
Electrode cleaning solution
Limescale
0.1 N HCl solution
English
13
Содержание sensIONTM+ pH31
Страница 2: ...English 3 Fran ais 19 Espa ol 37 Portugu s 54 72 86 102 117 2...
Страница 5: ...1 2 3 4 English 5...
Страница 15: ...1 2 English 15...
Страница 22: ...1 2 3 4 22 Fran ais...
Страница 33: ...1 2 Fran ais 33...
Страница 50: ...1 2 50 Espa ol...
Страница 57: ...1 2 3 4 Portugu s 57...
Страница 68: ...1 2 68 Portugu s...
Страница 73: ...2005 8 12 2002 98 EC sensION sensION PH31 pH ORP mV PC 1 1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 o 3 7 4 8 73...
Страница 74: ...1 2 3 4 74...
Страница 75: ...GFCI GFI 1 2 2 3 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 3 75...
Страница 82: ...1 2 82...
Страница 87: ...2005 8 12 EU 2002 98 EC sensION sensION PH31 pH ORP mV PC 1 1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 O 3 7 4 8 87...
Страница 88: ...1 2 3 4 88...
Страница 89: ...AC GFCI GFI AC 1 2 2 3 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 89...
Страница 90: ...3 AC 1 5 2 6 3 7 4 ID ID 90...
Страница 97: ...99 pH 1 1 pH 0 1 N HCl 97...
Страница 98: ...1 2 98...
Страница 103: ...2005 8 12 EU 2002 98 EC sensION sensION PH31 pH ORP mV PC 1 1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 3 7 4 8 103...
Страница 104: ...1 2 3 4 104...
Страница 105: ...AC GFCI GFI AC 1 2 2 3 AC 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 3 AC 105...
Страница 106: ...1 RETURN 5 DOWN 2 MEASUREMENT 6 ON OFF 3 UP 7 LEFT 4 RIGHT ID ID 4 1 ID 4 C F 2 pH ORP mV 5 3 AC 1 2 DATA OUTPUT 106...
Страница 112: ...114 pH 1 1 pH 0 1 N HCl 112...
Страница 113: ...1 2 113...
Страница 114: ...2 58 mV 58mV 112 a 70 t 100 112 2 112 112 3 pH 12 78 19 C pH 112 C 60 112 150 300 114...
Страница 118: ...12 2549 EU Directive 2002 98 EC sensION sensION PH31 pH ORP mV 1 1 1 5 2 pH 4 01 pH 7 00 pH 10 01 6 3 7 4 8 118...
Страница 119: ...1 2 119...
Страница 120: ...3 4 AC Ground Fault Circuit Interrupt GFCI GFI AC AC 1 2 2 3 AC 3 4 120...
Страница 121: ...2 1 5 DIN 2 6 3 RS 232 7 pH 4 8 3 AC 121...
Страница 122: ...1 RETURN 5 DOWN 2 MEASUREMENT 6 3 UP 7 LEFT 4 RIGHT ID ID 4 1 ID 4 C F 2 pH ORP mV 5 3 AC 1 2 DATA OUTPUT 122...
Страница 127: ...3 17 40 GLP 17 40 4 LabCom Easy 125 17 129 1 pH 1 pH 0 1 N HCl 127...
Страница 128: ...1 2 128...
Страница 132: ......
Страница 133: ......