background image

SYSTEM

SYSTEM

  Sample ID

   

  Data Output

  Type of report

  Sample ID

   

  Data Output

  Type of report

SYSTEM

Data Output

Deactivated

For printer

For computer

SYSTEM

Data Output

Deactivated

For printer

For computer

SYSTEM

Data Output

For computer

  Terminal

  ComLabo

  ComLabo Easy

10-04-2010     17:15
  

  

  SYSTEM

 

5. Sending data via the RS 232 C

6. Maintenance

Specifications

Baud Rate: 9600 bps
Word length: 8 bits
Stop bits: 2 bits
Parity: None

How to activate communication

At the start-up of the instrument or from “System”, see page 
16 .

Data sending to a printer 

(thermic or dot printer)

40 columns printers .
Thermic: Compatible EPSON (ESC/POS emulation)
Dot-Matrix: Compatible CITIZEN

Data sending to PC, Hyperterminal.

Data sending through the RS232C in printer format .
With the standard application Hyperteminal from Windows, the 
data can be captured in the same format as the printer reports, 
see pages 12 and 13 . They can be exported to an Excel program, 
as well .
See PC connection cables at page 20 (Replacement parts) .

Data sending to PC, ComLabo software

ComLabo 

(PN LZW8999.99). ComLabo

 Software allows control-

ling from a PC several modules, pH-meters, conductivity meters, 
automatic burettes, Sampler, etc .

ComLabo Easy

 

(PN LZW8997.99). ComLabo Easy

 Software allows 

obtaining data of pH-meters and conductivity meters from a PC .

Main features shared

Windows based software .
Data Logger . All calibration and measuring data are stored in 
the PC .
Sample code assignment from PC
Data export to EXCEL .

6.1. Clean the instrument 

Important note:

 Never use cleaning such as turpentine, acetone 

or similar products to clean the instrument including the display 
and accessories .
Only clean the housing and accessories using a soft, damp cloth . 
Mild soap solution may also be used . Dry the cleaned parts care-
fully with a soft cotton cloth .

 

 CAUTION

    

When using chemicals or solvents, comply with the instruc-
tions of the producer and all local safety regulations .

6.2. Sensor cleaning

See the manual of the sensor in use .

19

ENGLISH

Содержание sension ph31

Страница 1: ...ACH Company 2010 All rights reserved Printed in Spain Manual Manual Manual de instru es Manuel d utilisateur 07 2010 Edition 1 07 2010 Edici n 1 07 2010 Edi o 1 07 2010 Edition 1 2010 7 DOC022 97 9024...

Страница 2: ......

Страница 3: ...ENGLISH ESPA OL FRAN AIS PORTUGU S CONTACT...

Страница 4: ......

Страница 5: ...iagram 7 4 5 pH calibration 8 4 6 Other options for calibration 9 4 7 pH measurements 10 4 8 Re programming pH measurement 11 4 9 Printing data of pH measurements 12 4 10 mV measurements 14 4 11 Data...

Страница 6: ...5 mV calibration Automatic recognition of redox standard 220 mV at 25 C 77 F Calibration at any mV value Manual introduction of the calibration parameters Temperature readjustment Correction of the te...

Страница 7: ...abels Read carefully all labels and tags attached to the instrument Accessories Buffer solutions 135 ml flasks Magnetic stirrer Calibration flasks with magnetic bar inside Electrolyte for the electrod...

Страница 8: ...ture 3 Drive the stick into the hole 4 Turn the shift in a counterclockwise until the stick is well fixed Place the o ring at 135 140 mm from the base in order to avoid the shock of the sensor against...

Страница 9: ...Make sure the power adapter matches your local power supply 4 1 Description Keys Switch on Switch off Confirms the option displayed When there is more than one option accepts the one marked in negati...

Страница 10: ...majority of applications pH Resolution 0 01 pH Isopotential pH 7 00 Measurements By stability with standard criterion In continuous pressing twice the key Calibration in 1 2 or 3 points with technica...

Страница 11: ...ution 0 01 Meas By stability Display mV No Limits No Isopotential pH 7 Measurement condit Technical Buffers DIN 19266 Buffers User buffers To a X value Data Introduction Theoretical cal Cal frequency...

Страница 12: ...values similar to those of the buffer solution It corrects the asymmetry potential of the electrode Two point calibration This is the most common calibration It is recommended to start with pH 7 as fi...

Страница 13: ...C 77 F Buffers DIN 19266 pH 1 679 4 006 6 865 7 000 9 180 10 012 and 12 454 User s buffers Suitable when the two above mentioned options will not be used ThesensION PH31allowsintroducingthetablepH Tem...

Страница 14: ...us 10 04 2010 17 15 pH MEASURE New calibration in 1 day 4 Operation 4 7 pH measurements Measurement by stability This is the default measuring mode for the sensION PH31 Meanwhile the reading is in evo...

Страница 15: ...limit pH 8 20 Start Measurement by stability 10 04 2010 17 15 pH MEASURE New calibration in 1 day Default programming standard Other options Resolution 0 01 pH 0 1 pH 0 01 pH 0 001 pH Measuring mode...

Страница 16: ...ry 24 h SYSTEM SYSTEM Sample ID Data Output Date Time Language Display contrast Sample ID Data Output Date Time Language Display contrast SYSTEM Data output Deactivated For printer For computer SYSTEM...

Страница 17: ...50 14T Serial number 7890 pHiso pH 7 pH 7 00 CALIBRATED 14 02 2010 08 50 37 Technical Sens A P T t St Buffers mV C s 4 01 7 00 100 3 0 1 25 6 25 CONDITIONS CHANNEL 1 STABILITY MEASUREMENTS STANDARD C...

Страница 18: ...ing Data Data introduction Compensation 000 0 mV Compensation 0 mV Continue Factory Adjust Use standard 220 mV Continue CALIBRATING mV To a X value Use specific standard Continue CALIBRATING mV 10 04...

Страница 19: ...ple pH C Time St 15 03 07 14 36 38510 10 15 23 5 00 15 30 15 03 07 14 40 38511 9 85 23 2 00 22 30 15 03 07 14 46 38512 10 23 22 8 00 18 30 16 03 07 08 36 Pozo1 7 48 23 1 00 25 30 pH CALIBRATION DATA C...

Страница 20: ...veral One sample SYSTEM Type of report Reduced Standard GLP SYSTEM Users Yes No SYSTEM Correction 25 0 C 0 0 C 85 0 C 0 0 C SYSTEM SYSTEM Display contrast Display contrast SYSTEM Display contrast Read...

Страница 21: ...Continue Calibration pH 4 Operation 4 13 Temperature readjustment The sensION PH31 allows the correction of the deviation presented by a temperature probe integrated in the electrode or separated sens...

Страница 22: ...temperature compensation in the measurements Example for electrode behavior a Electrode calibration line at room temperature b A calibration line with pHiso 7 to apply when measure at 70 C c Real cal...

Страница 23: ...y can be exported to an Excel program as well See PC connection cables at page 20 Replacement parts Data sending to PC ComLabo software ComLabo PN LZW8999 99 ComLabo Software allows control ling from...

Страница 24: ...perature Check the electrode MEASURING by stability UNSTABLE READING Time 300 s Check temperature Check the electrode MEASURING by stability pH out of range Check the electrode MEASURING by stability...

Страница 25: ...repair or attempted repair not authorized by Hach Company Any product not used in accordance with the instructions furnished by Hach Company Freight charges to return merchandise to Hach Company Freig...

Страница 26: ......

Страница 27: ...o hach lange nl www hach lange nl HACH LANGE APS kandevej 21 DK 2700 Br nsh j Tel 45 36 77 29 11 Fax 45 36 77 49 11 info hach lange dk www hach lange dk HACH LANGE AB Vinthundsv gen 159A SE 128 62 Sk...

Страница 28: ......

Отзывы: