4. Operation

4.16. Isopotential (pHiso)

Definition

Potential (mV) of one pH electrode that does not alter with the 
temperature . This is the intersection point of the different cali-
bration lines at different calibration temperature . Normally, this 
value corresponds to pH 7 . Nevertheless, in reality the value is 
slightly different .

Application

It is recommended to determine the pH of the real isopotential 
(

 7) when after calibration at room temperature, the measure-

ments will be performed at different temperature, and in addi-
tion a very high precision is required .

sensION

+

 

MM340

The instrument offers three possibilities for the application of pHiso:

pH 7

: A value generally accepted by all the manufacturers .

Introducing the real value:

 If the pHiso is known, it can be 

introduced by the user .

Value calculation

, by the instrument .

Procedure:

- Calibrate the instrument in 2 or 3 points with the buffers at 

room temperature, following the common procedure .

- Heat the buffers . Their temperature must differ with 10 or more 

degree centigrade from the calibration temperature .

- Select “Calculate” from the menu “Measuring conditions” (see 

pH measurement re-programming on page 11) and follow the 
instructions of the pH-meter .

- The instrument will calculate the pHiso of the electrode in use 

and will memorize and apply it in the corresponding calculation 
for the temperature compensation in the measurements .

Example for electrode behavior

a)

  Electrode calibration line at room temperature

b)

  A calibration line with pHiso=7 to apply when measure at 

70ºC (158 °F)

c)

  Real calibration line of the electrode at 70ºC (158 °F)

P1) 

Classic isopotential, pHiso 7

P2)

 Real isopotential of the electrode, pHiso 6 .

Error:

 Appling the classic isopotential (pHiso7) the pH measure-

ment in buffer 5 at 70ºC (122 to 158 °F) will be 5 .15 .

4.17. Recognized buffers

Buffer solutions

The 

sensION

+

 

MM340

 has in memory the following data (pH 

values as a function of the temperature) .

Technical buffers (DIN 19267)

Buffer solutions DIN 19266

19

ENGLISH

ºC

ºF

pH

mV

0

32

2 .01

4 .01

7 .12

9 .52

11 .45

--

10

50

2 .01

4 .00

7 .06

9 .38

11 .20

245

20

68

2 .00

4 .00

7 .02

9 .26

11 .00

228

25

77

2.00

4.01

7.00

9.21 10.90 220

30

86

2 .00

4 .01

6 .99

9 .16

10 .81

212

40

104

2 .00

4 .03

6 .97

9 .06

10 .64

195

50

122

2 .00

4 .06

6 .97

8 .99

10 .48

178

60

140

2 .00

4 .10

6 .98

8 .93

10 .23

160

70

158

2 .01

4 .16

7 .00

8 .88

10 .19

142

80

176

2 .01

4 .22

7 .04

8 .83

10 .06

--

90

194

2 .01

4 .30

7 .09

8 .79

9 .93

--

ºC

ºF

pH

5

32

1 .668 4 .004 6 .951 7 .087 9 .395 10 .245 13 .207

10

50

1 .670 4 .000 6 .923 7 .059 9 .332 10 .179 13 .003

20

68

1 .675 4 .001 6 .881 7 .016 9 .225 10 .062 12 .627

25

77 1.679 4.006 6.865 7.000 9.180 10.012 12.454

30

86

1 .683 4 .012 6 .853 6 .987 9 .139 9 .966 12 .289

40

104 1 .694 4 .031 6 .838 6 .970 9 .068 9 .889 11 .984

50

122 1 .707 4 .057 6 .833 6 .964 9 .011 9 .828 11 .705

60

140 1 .723 4 .085 6 .836 6 .968 8 .962

--

11 .449

70

158 1 .743 4 .126 6 .845 6 .982 8 .921

--

--

80

176 1 .766 4 .164 6 .859 7 .004 8 .885

--

--

90

194 1 .792 4 .205 6 .877 7 .034 8 .850

--

--

a

c

b

pH 5.00

pH 5.15

P2

P1

Содержание sensION MM340

Страница 1: ...ACH Company 2010 All rights reserved Printed in Spain Manual Manual Manual de instru es Manuel d utilisateur 07 2010 Edition 1 07 2010 Edici n 1 07 2010 Edi o 1 07 2010 Edition 1 2010 7 DOC022 97 9025...

Страница 2: ...libration 8 4 6 Other options for calibration 9 4 7 pH measurements 10 4 8 Re programming pH measurement 11 4 9 Printing data of pH measurements 12 4 10 mV measurements 14 4 11 Ion selective measureme...

Страница 3: ...7 F Calibration at any mV value Manual introduction of the calibration parameters ISE calibration 2 to 5 programmable standards Selectable calibration units mol l M g l mg l ppm mmol l mM mol l or Up...

Страница 4: ...d lead to damage to the instrument Important information that requires special emphasis Note Information that supplements points in the main text 2 1 2 Precautionary labels Read carefully all labels a...

Страница 5: ...ure 3 Drive the stick into the hole 4 Turn the shift in a counterclockwise until the stick is well fixed Place the o ring at 135 140 mm from the base in order to avoid the shock of the sensor against...

Страница 6: ...ctor 11 Plug in power supply 12 V 3 2 Connections DANGER Make sure the power adapter matches your local power supply 4 1 Description Keys Switch on Switch off Confirms the option displayed When there...

Страница 7: ...ution 0 01 pH Measurements By stability with standard criterion In continuous pressing twice the key Calibration in 1 2 or 3 points with technical buffers Calibration frequency every 24 hours Channel...

Страница 8: ...DIN 19266 buffers User buffers To a X value Data introduction Theoretical cal Cal frequency Calibration pH Start calibration Channel 1 Technical bufferes Consult Print Erase DATA LOGGER C1 Start measu...

Страница 9: ...en measuring pH values similar to those of the buffer solution It corrects the asymmetry potential of the electrode Two point calibration This is the most common calibration It is recommended to start...

Страница 10: ...06 6 865 7 000 9 180 10 012 and 12 454 User s buffers Suitable when the two above mentioned options will not be used The sensION MM340 allows introducing the table pH Temp for the desired buffers See...

Страница 11: ...01 MEASURING continuous Channel 1 pH MEASURE New CAL in 12 h 10 04 2010 17 15 4 Operation 4 7 pH measurements Measurement by stability This is the default measuring mode for the sensION MM340 Meanwhi...

Страница 12: ...art Measurement Channel 1 by stability 10 04 2010 17 15 Channel 1 pH MEASURE New CAL in 12 h Default programming standard Other options Resolution 0 01 pH 0 1 pH 0 01 pH 0 001 pH Measuring mode By sta...

Страница 13: ...ion frequency 24 h SYSTEM SYSTEM Measurement units Sample ID Data output Measurement units Sample ID Data output SYSTEM Data output Deactivated For printer For computer SYSTEM Data output Deactivated...

Страница 14: ...ial number 7890 pHiso pH 7 pH 7 00 CALIBRATED 14 02 2010 08 50 37 Technical Sens A P T t St Buffers mV C s 4 01 7 00 100 3 0 1 25 6 25 CONDITIONS CHANNEL 1 STABILITY MEASUREMENTS STANDARD CRITERIA Sam...

Страница 15: ...hannel 2 ISE Data Time 04 06 2010 09 30 Monday SYSTEM Codification Automatic Manual SYSTEM Measurement units Sample ID Data output SYSTEM Measurement units Channel 1 pH Channel 2 ISE Save changes SYST...

Страница 16: ...CALIBRATION 10 04 2010 17 15 4 Operation 4 11 Measurement with ion selective electrode Curve number 1 sensION MM340 can store in memory up to 4 calibration curves for each channel Measurement by stab...

Страница 17: ...Minimum value pH 13 75 1 15 mV T C 38 2 14 3 MEASUREMENT DATA Channel 1 SN 70602 V1 0 Monday 30 July 2010 15 50 Sample pH C Time St 15 03 07 14 36 38510 10 15 23 5 00 15 30 15 03 07 14 40 38511 9 85...

Страница 18: ...Display contrast SYSTEM Stirring Stirrer N 1 OFF 25 ON SYSTEM Stir Stirrer N 1 ON 25 OFF SYSTEM Readjust temp Stirring Temperature units SYSTEM Temperature units Celsius Fahrenheit SYSTEM ENGLISH GER...

Страница 19: ...bration pH 4 Operation 4 14 Temperature readjustment The sensION MM340 allows the correction of the deviation presented by a temperature probe integrated in the electrode or separated sensor at 25 and...

Страница 20: ...compensation in the measurements Example for electrode behavior a Electrode calibration line at room temperature b A calibration line with pHiso 7 to apply when measure at 70 C 158 F c Real calibratio...

Страница 21: ...the printer reports see pages 12 and 13 They can be exported to an Excel program as well See PC connection cables at page 21 Replacement parts Data sending to PC ComLabo software ComLabo PN LZW8999 99...

Страница 22: ...00 s Check temperature Check the electrode MEASURING by stability pH out of range Check the electrode MEASURING by stability out of range C Check temp sensor MEASURING by stability MEASURE STABLE pH 1...

Страница 23: ...any repair or attempted repair not authorized by Hach Company Any product not used in accordance with the instructions furnished by Hach Company Freight charges to return merchandise to Hach Company F...

Страница 24: ...o hach lange nl www hach lange nl HACH LANGE APS kandevej 21 DK 2700 Br nsh j Tel 45 36 77 29 11 Fax 45 36 77 49 11 info hach lange dk www hach lange dk HACH LANGE AB Vinthundsv gen 159A SE 128 62 Sk...

Отзывы: