Hach sensION MM340 Скачать руководство пользователя страница 10

Adjust value

pH 3.09 

pH 3.08          25˚C

00:06

CALIBRATING pH 

CALIBRATING pH 

  ...

  

  Start calibration

  Channel 1  

Technical buffers

Calibration pH 

Theoretical cal.

Pot. asym.: 0 mV

Slope: 59.16mV/pH

Temperature: 25 °C

Continue...

Technical buffers

Use first buffer

Continue...

CALIBRATING pH 

DIN 19266 buffers

Use First Buffer

Continue...

CALIBRATING pH 

To a X value

Use standard

Continue...

CALIBRATING pH 

Type of 

calibration

  Technical buffers

  DIN 19266 buffers

  User Buffers

  To a X value

Calibration pH 

Type of 

calibration

  Technical buffers

  DIN 19266 buffers

  User Buffers

  To a X value

Calibration pH 

Type of 

calibration

  Technical buffers

  DIN 19266 buffers

  User Buffers

  To a X value

Calibration pH 

Type of 

calibration

  Technical buffers

  DIN 19266 buffers

  User Buffers

  To a X value

Calibration pH 

Type of 

calibration

  Data introduction

  Theoretical cal.

Calibration pH  

Type of 

calibration

  Data introduction

  Theoretical cal.

Calibration pH 

Channel 1

  

pH CALIBRATION

  

10-04-2010   17:15

Stability c: standard

C

alibration type

Frequency

View mV

Calibration pH

Stability c: standard

Calibration type

Frequency

View mV

Calibration pH

User buffers

1st buffer pH

Continue...

CALIBRATING pH

User buffers

Start calibration

No. Buffers:  1

Buffer 1

CALIBRATING pH

Calibration pH

Calibration frequency 

Automatic warning. 

1 day  

00

 hours

Calibration pH

Introducing data

Pot. asym.:

 0

 mV

Slope: 59.16mV/pH

Temperature: 25 °C

Save changes

4. Operation

4.5. Other options for calibration

Automatic buffers recognition

Technical buffers

pH 2 .00, 4 .01, 7 .00, 9 .21 and 10 .90 at 25ºC (77ºF) .

Buffers DIN 19266

pH 1 .679, 4 .006, 6 .865, 7 .000, 9 .180, 10 .012 and 12 .454 .

User’s buffers 

Suitable  when  the  two  above  mentioned  options  will  not  be 
used .  The 

  MM340

  allows  introducing  the 

table pH/Temp . for the desired buffers . See System, page 17 .

Calibration to X value

Manual re-adjustment of the measured pH at any value of the 
scale . The pH readjustment is equivalent to calibration with only 
one buffer .

Calibration data introduction

Manual  introduction  of  calibration  data  for  an  electrode  cali-
brated previously, for example in calibration laboratory .

Theoretical calibration

Selecting “theoretical calibration” the calibration data of the elec-
trode in use are replaced by the theoretical values at 25ºC (77ºF) .
This step is prior to the obtention of a calibration certificate .

Calibration frequency

Select  between  0  h  and  7  days .  The  instrument  by  default  is 
programmed  for  daily  calibration  and  indicates  the  remaining 
time to the new calibration . 
Selecting 0 h the automatic re-calibration warning is deactivated .

Stability criteria. 

Rapid criterion (variation < 0 .02 pH in 6 s) .
Standard criterion (variation < 0 .01 pH in 6 s) .
Strict criterion (variation < 0 .002 pH in 6 s) .
It is highly recommended to apply the same stability criteria to 
measure and calibrate .

Important note:

Once the calibration is 
re-programmed, it will be 
performed with the new 
programming .

9

ENGLISH

Start of calibration 
with automatic 
recognition of the 
selected buffers.

Start calibration

The instrument goes 
to MEASURE menu. 
The theoretical cali-
bration will be used.

The instrument goes to the 
option “CALIBRATE pH” and 
it must be recalibrated.

The instrument goes to  
MEASURE menu. The 
introduced calibration 
data will be used.

See Stability Criteria  
and View mV  
on page 10.

The instrument 
goes to Measure.

See page 17.

Содержание sensION MM340

Страница 1: ...ACH Company 2010 All rights reserved Printed in Spain Manual Manual Manual de instru es Manuel d utilisateur 07 2010 Edition 1 07 2010 Edici n 1 07 2010 Edi o 1 07 2010 Edition 1 2010 7 DOC022 97 9025...

Страница 2: ...libration 8 4 6 Other options for calibration 9 4 7 pH measurements 10 4 8 Re programming pH measurement 11 4 9 Printing data of pH measurements 12 4 10 mV measurements 14 4 11 Ion selective measureme...

Страница 3: ...7 F Calibration at any mV value Manual introduction of the calibration parameters ISE calibration 2 to 5 programmable standards Selectable calibration units mol l M g l mg l ppm mmol l mM mol l or Up...

Страница 4: ...d lead to damage to the instrument Important information that requires special emphasis Note Information that supplements points in the main text 2 1 2 Precautionary labels Read carefully all labels a...

Страница 5: ...ure 3 Drive the stick into the hole 4 Turn the shift in a counterclockwise until the stick is well fixed Place the o ring at 135 140 mm from the base in order to avoid the shock of the sensor against...

Страница 6: ...ctor 11 Plug in power supply 12 V 3 2 Connections DANGER Make sure the power adapter matches your local power supply 4 1 Description Keys Switch on Switch off Confirms the option displayed When there...

Страница 7: ...ution 0 01 pH Measurements By stability with standard criterion In continuous pressing twice the key Calibration in 1 2 or 3 points with technical buffers Calibration frequency every 24 hours Channel...

Страница 8: ...DIN 19266 buffers User buffers To a X value Data introduction Theoretical cal Cal frequency Calibration pH Start calibration Channel 1 Technical bufferes Consult Print Erase DATA LOGGER C1 Start measu...

Страница 9: ...en measuring pH values similar to those of the buffer solution It corrects the asymmetry potential of the electrode Two point calibration This is the most common calibration It is recommended to start...

Страница 10: ...06 6 865 7 000 9 180 10 012 and 12 454 User s buffers Suitable when the two above mentioned options will not be used The sensION MM340 allows introducing the table pH Temp for the desired buffers See...

Страница 11: ...01 MEASURING continuous Channel 1 pH MEASURE New CAL in 12 h 10 04 2010 17 15 4 Operation 4 7 pH measurements Measurement by stability This is the default measuring mode for the sensION MM340 Meanwhi...

Страница 12: ...art Measurement Channel 1 by stability 10 04 2010 17 15 Channel 1 pH MEASURE New CAL in 12 h Default programming standard Other options Resolution 0 01 pH 0 1 pH 0 01 pH 0 001 pH Measuring mode By sta...

Страница 13: ...ion frequency 24 h SYSTEM SYSTEM Measurement units Sample ID Data output Measurement units Sample ID Data output SYSTEM Data output Deactivated For printer For computer SYSTEM Data output Deactivated...

Страница 14: ...ial number 7890 pHiso pH 7 pH 7 00 CALIBRATED 14 02 2010 08 50 37 Technical Sens A P T t St Buffers mV C s 4 01 7 00 100 3 0 1 25 6 25 CONDITIONS CHANNEL 1 STABILITY MEASUREMENTS STANDARD CRITERIA Sam...

Страница 15: ...hannel 2 ISE Data Time 04 06 2010 09 30 Monday SYSTEM Codification Automatic Manual SYSTEM Measurement units Sample ID Data output SYSTEM Measurement units Channel 1 pH Channel 2 ISE Save changes SYST...

Страница 16: ...CALIBRATION 10 04 2010 17 15 4 Operation 4 11 Measurement with ion selective electrode Curve number 1 sensION MM340 can store in memory up to 4 calibration curves for each channel Measurement by stab...

Страница 17: ...Minimum value pH 13 75 1 15 mV T C 38 2 14 3 MEASUREMENT DATA Channel 1 SN 70602 V1 0 Monday 30 July 2010 15 50 Sample pH C Time St 15 03 07 14 36 38510 10 15 23 5 00 15 30 15 03 07 14 40 38511 9 85...

Страница 18: ...Display contrast SYSTEM Stirring Stirrer N 1 OFF 25 ON SYSTEM Stir Stirrer N 1 ON 25 OFF SYSTEM Readjust temp Stirring Temperature units SYSTEM Temperature units Celsius Fahrenheit SYSTEM ENGLISH GER...

Страница 19: ...bration pH 4 Operation 4 14 Temperature readjustment The sensION MM340 allows the correction of the deviation presented by a temperature probe integrated in the electrode or separated sensor at 25 and...

Страница 20: ...compensation in the measurements Example for electrode behavior a Electrode calibration line at room temperature b A calibration line with pHiso 7 to apply when measure at 70 C 158 F c Real calibratio...

Страница 21: ...the printer reports see pages 12 and 13 They can be exported to an Excel program as well See PC connection cables at page 21 Replacement parts Data sending to PC ComLabo software ComLabo PN LZW8999 99...

Страница 22: ...00 s Check temperature Check the electrode MEASURING by stability pH out of range Check the electrode MEASURING by stability out of range C Check temp sensor MEASURING by stability MEASURE STABLE pH 1...

Страница 23: ...any repair or attempted repair not authorized by Hach Company Any product not used in accordance with the instructions furnished by Hach Company Freight charges to return merchandise to Hach Company F...

Страница 24: ...o hach lange nl www hach lange nl HACH LANGE APS kandevej 21 DK 2700 Br nsh j Tel 45 36 77 29 11 Fax 45 36 77 49 11 info hach lange dk www hach lange dk HACH LANGE AB Vinthundsv gen 159A SE 128 62 Sk...

Отзывы: