Figure 9 Connect conduit or a power cord
1
AC power terminal
3
Protective earth ground
2
Cables limit: do not put cables above the line.
4
Conduit hub (or strain relief fitting for power cord)
Table 3 Wiring information—AC power
Terminal
Description
Color—North America
Color—EU
L
Hot (L1)
Black
Brown
N
Neutral (N)
White
Blue
Protective earth ground
Green
Green with yellow stripe
Connect the high-voltage relays
D A N G E R
Electrocution hazard. Always remove power to the instrument before making electrical connections.
W A R N I N G
Potential Electrocution Hazard. Power and relay terminals are designed for only single wire termination.
Do not use more than one wire in each terminal.
W A R N I N G
Potential fire hazard. Do not daisy-chain the common relay connections or jumper wire from the mains
power connection inside the instrument.
16
English
Содержание SC4200c
Страница 2: ...English 3 Espa ol 23 Fran ais 45 67 2...
Страница 68: ...ESD 68...
Страница 71: ...SC4200c 1 MSM Web MSM WiFi LAN MSM MSM 1 1 6 USB 2 7 USB 3 SIM 3 8 4 4 9 5 10 3 SIM 4 71...
Страница 73: ...IT No reply hach com donotreply hach com Hach 2000 m 6562 ft 2000 m 16 cm 6 30 in 4 3 72 73...
Страница 74: ...3 19 65 mm 0 75 2 5 in 4 72 74...
Страница 75: ...4 5 15 mm 75...
Страница 76: ...5 6 6 1 5 SC 1 1 2 6 SC 2 2 3 TCP IP 7 7 9 4 LAN 8 ESD ESD 7 8 9 76...
Страница 77: ...7 7 8 77...
Страница 78: ...8 78...
Страница 79: ...3 m 10 ft 0 75 2 5 mm2 18 12 AWG 3 m 10 ft 70 C 158 F 0 75 mm2 18 AWG 9 3 PCBA PCBA 9 79...
Страница 80: ...9 1 3 2 4 3 L L1 N N 80...
Страница 82: ...10 4 1 1 NC 4 2 NC 2 1 5 2 3 1 NO 6 2 NO NC NO Profibus PCBA 8 78 1 4 Nm 12 4 in lb 7 77 82...
Страница 83: ...SC SC 6 76 11 83 11 12 12 1 4 2 2 1 5 2 3 1 5 83...
Страница 85: ...1 2 MSM MSM MSM mA 15 15 85...
Страница 86: ...MSM Web SIM APN Web SIM SIM LAN 6 76 Web MSM 86...
Страница 87: ......