Problem
Possible cause
Solution
The measurements take too
much time to complete and/or
the results are consistently too
low.
The UV reactor is almost
expired.
Replace the UV reactor. Make
sure that there is a sufficient
quantity of carrier gas.
The fittings around the reactor
are loose.
Make sure that the fittings
around the reactor are tight.
The UV reactor does not work.
The UV reactor is expired or
damaged.
Replace the UV reactor.
Contact technical support.
The sample is not found.
There is not enough solution in
the reagent bottle.
Fill the reagent bottle with
solution.
The sample tube is not
connected to the auto sampler
or it is not in the sample flask.
Make sure that the sample tube
is connected to the auto
sampler or that it is in the
sample flask.
A hardware failure.
Make sure that there is a
sample to use for the
procedure. Contact technical
support.
Significant carryover occurs
from sample to sample.
The reactor is not fully flushed
after each measurement.
Examine the carrier gas supply.
If the problem continues,
contact technical support.
A pressure error frequently
shows during the system health
test.
The carrier gas is empty or
low.
Measurements frequently show
the result as "0".
The background TOC reading
is larger than the sample TOC
reading.
Fill the reagent bottle with
water, then do a new
background qualification
procedure.
The report output contains no
information.
An error occurred when a
report was made.
Contact technical support.
Changes are not saved when
the instrument power is set to off
then on.
An error occurred when a
setting was changed.
The qualification procedure
does not start.
An error occurred during a
qualification procedure.
The USB import and/or export
function does not operate with a
USB flash drive.
The USB flash drive is not
connected.
Connect the USB flash drive to
the instrument.
The USB flash drive is not
found or is corrupt.
Fix or format the USB flash
drive (if necessary), then
connect it to the instrument.
The network connection does
not show.
The ethernet cable is not
connected.
Make sure that the ethernet
cables is connected to the
instrument.
English
21
Содержание QbD1200+
Страница 1: ...DOC022 97 80647 QbD1200 09 2021 Edition 1 User Manual...
Страница 2: ...Table of Contents English 3 24 44 65 86 106 2...
Страница 27: ...3 1 4 3 1 5 MSDS SDS 3 1 6 Canadian Radio Interference Causing Equipment Regulation ICES 003 A 27...
Страница 29: ...3 1 4 2 5 3 6 GLS 4 4 1 MSDS SDS 29...
Страница 30: ...4 2 4 2 1 GLS GLS GLS 30...
Страница 31: ...4 2 2 4 2 3 31...
Страница 32: ...4 2 4 1 241 689 kPA 35 100 psi 2 1 8 3 1 24 4 5 4 3 4 3 1 USB 1 3 9 84 24 USB USB 4 3 2 4 3 3 5 5 1 4 1 32...
Страница 33: ...4 1 5 2 6 3 1 7 4 8 1 USB 1 33...
Страница 34: ...1 SST500 SDBS SST8000 6 6 1 1 2 3 password 4 5 6 35 34...
Страница 36: ...3 7 1 3 38 1 2 3 7 1 4 1 2 1 64 1 5 ID ID 36...
Страница 37: ...4 ppm 2 mL 5 ppm ppb mg L g L psi mbar TIC TIC ppb ppb 3 7 1 5 1 2 AS0640 7 2 7 2 1 1 2 38 USB ID 37...
Страница 38: ...3 4 7 2 2 1 2 1 5 ID ID ID 3 4 5 6 7 7 3 7 3 1 1 2 PDF CSV XML USB 37 USB USB 7 4 MSDS SDS 38...
Страница 39: ...1 2 SST500 SDBS SST8000 3 ID 2 4 8 MSDS SDS 8 1 1 2 8 2 2 39...
Страница 40: ...8 3 1 a 30 mL b 30 mL c 2 3 30 8 4 1 2 8 5 USB 1 2 USB USB 3 setup exe 4 5 8 6 1 2 3 GLS 24 4 40...
Страница 41: ...5 GLS 6 7 9 0 TOC TOC USB USB USB USB USB USB USB USB USB USB 41...
Страница 43: ...USP 8 ppm 9459800 9459900 9460000 9460100 43...
Страница 47: ...3 1 3 3 1 4 3 1 5 MSDS SDS 47...
Страница 48: ...3 1 6 ICES 003 A A FCC PART 15 A FCC 15 1 2 FCC 15 A 1 2 3 4 5 3 2 3 3 m 9 8 48...
Страница 49: ...3 1 4 2 5 3 6 GLS 4 4 1 MSDS SDS 49...
Страница 50: ...4 2 4 2 1 GLS GLS GLS 50...
Страница 51: ...4 2 2 4 2 3 51...
Страница 52: ...4 2 4 1 241 689 kPA 2 3 175 mm 3 1 44 4 5 4 3 4 3 1 3 USB 1 3 44 USB USB 4 3 2 4 3 3 5 5 1 4 1 52...
Страница 53: ...4 1 5 2 6 3 1 7 4 8 1 USB 1 53...
Страница 54: ...1 SST500 SDBS SST8000 6 6 1 1 2 admin 3 password 4 5 6 55 54...
Страница 55: ...7 7 1 7 1 1 USB USB 1 2 2 5 10 30 60 USB 3 7 1 2 1 2 1 3 5 10 30 60 1 15 20 30 45 60 90 120 55...
Страница 56: ...1 2 3 4 5 5 10 30 60 3 7 1 3 59 1 2 3 7 1 4 1 2 1 64 1 5 ID ID 56...
Страница 57: ...4 ppm 2 mL 5 UV ppm ppb mg L g L SI psi mbar TIC TIC ppb ppb 3 3 7 1 5 1 2 AS0640 Ok 57...
Страница 58: ...7 2 7 2 1 1 2 58 USB 1 ID 3 4 7 2 2 1 2 1 5 ID ID ID 3 4 5 6 7 58...
Страница 59: ...7 3 7 3 1 1 2 PDF CSV XML USB 57 USB USB 7 4 MSDS SDS 1 2 SST500 SDBS SST8000 3 ID ID 2 4 2 59...
Страница 60: ...8 MSDS SDS 8 1 1 2 8 2 8 3 1 1 a 30 mL b 30 mL c 2 3 30 1 8 4 UV 60...
Страница 61: ...1 2 UV UV UV 8 5 Web Web USB 1 2 USB USB 3 setup exe 4 5 8 6 1 2 3 GLS 44 4 5 GLS 6 TOC 7 9 61...
Страница 62: ...UV UV UV UV UV 0 TOC TOC USB USB USB USB USB USB 62...
Страница 63: ...USB USB USB USB Invalid Password Access Denied 10 Web UV 9880200 9891700 9898900 63...
Страница 68: ...3 1 3 3 1 4 3 1 5 UV UV UV MSDS SDS 68...
Страница 70: ...3 1 4 2 5 3 6 GLS 4 4 1 MSDS SDS 70...
Страница 71: ...4 2 4 2 1 GLS GLS GLS 71...
Страница 72: ...4 2 2 4 2 3 72...
Страница 73: ...4 2 4 1 241 689 kPA 35 100 psi 2 3 1 65 4 5 4 3 4 3 1 USB 3 1 3 9 84 ft 65 USB USB 4 3 2 4 3 3 5 5 1 4 1 73...
Страница 74: ...4 1 5 2 6 3 1 7 4 8 1 USB 1 74...
Страница 75: ...1 SST500 SDBS SST8000 6 6 1 1 2 3 4 5 6 76 75...
Страница 76: ...7 7 1 7 1 1 USB USB 1 2 2 5 10 30 60 USB 3 7 1 2 1 2 1 3 5 10 30 60 1 15 20 30 45 60 90 120 76...
Страница 77: ...1 2 3 4 5 5 10 30 60 3 7 1 3 80 1 2 3 7 1 4 1 2 1 64 1 5 ID ID 77...
Страница 78: ...4ppm 2mL 5 UV ppm ppb mg L g L SI kPa mbar TIC TIC ppb ppb 3 7 1 5 1 2 AS0640 78...
Страница 79: ...7 2 7 2 1 1 2 79 USB 1 ID 3 4 7 2 2 1 2 1 5 ID ID ID 3 4 5 6 7 79...
Страница 80: ...7 3 7 3 1 1 2 PDF CSV XML USB 78 USB USB 7 4 MSDS SDS 1 2 SST500 SDBS SST8000 3 ID 2 4 2 80...
Страница 81: ...8 MSDS SDS 8 1 1 2 8 2 8 3 1 a 30mL b 100mL c 2 3 30 8 4 81...
Страница 82: ...1 2 UV UV UV 8 5 USB 1 2 USB USB 3 setup exe 4 5 8 6 1 2 3 GLS 65 4 5 GLS 6 7 9 82...
Страница 83: ...0 TOC TOC USB USB USB USB USB USB 83...
Страница 84: ...USB USB USB USB 10 9880200 9891700 9898900 9890500 84...
Страница 89: ...3 1 3 3 1 4 3 1 5 UV UV UV 89...
Страница 91: ...3 1 4 2 5 3 6 GLS 4 4 1 MSDS SDS 91...
Страница 92: ...4 2 4 2 1 GLS GLS GLS 92...
Страница 93: ...4 2 2 4 2 3 93...
Страница 94: ...4 2 4 1 241 689 kPA 35 100 psi 2 2 3 1 86 4 5 4 3 4 3 1 USB 3 1 3 9 84 86 USB USB 4 3 2 4 3 3 5 5 1 4 1 94...
Страница 95: ...4 1 5 2 6 3 1 7 4 8 1 USB 1 95...
Страница 96: ...1 SST500 SDBS SST8000 6 6 1 1 2 admin 3 password 4 5 6 97 96...
Страница 98: ...3 7 1 3 100 1 2 3 7 1 4 1 2 1 64 1 5 ID ID 98...
Страница 99: ...4 ppm 2 5 UV ppm ppb SI g psi mbar TIC TIC ppb ppb 3 3 7 1 5 1 2 1 AS0640 7 2 7 2 1 1 2 100 USB 99...
Страница 100: ...ID 3 4 7 2 2 1 2 1 5 ID ID ID 3 4 5 6 7 7 3 7 3 1 1 2 PDF CSV XML USB 99 USB USB 100...
Страница 101: ...7 4 MSDS SDS 1 2 SST500 SDBS SST8000 3 ID 2 4 8 MSDS SDS 8 1 1 2 2 101...
Страница 102: ...8 2 8 3 1 1 a 30 b 30 c 2 3 30 8 4 UV 1 2 UV UV UV 8 5 USB 1 2 USB USB 3 setup exe 4 5 102...
Страница 103: ...8 6 1 2 3 GLS 86 4 5 GLS 6 7 9 UV UV UV UV UV 0 TOC TOC 103...
Страница 104: ...USB USB USB USB USB USB USB USB USB opto isolator USB 10 104...
Страница 109: ...3 1 4 3 1 5 UV UV UV MSDS SDS 3 1 6 Canadian Radio Interference Causing Equipment Regulation ICES 003 Class A 109...
Страница 111: ...3 1 4 2 5 3 6 GLS 4 4 1 MSDS SDS 111...
Страница 112: ...4 2 4 2 1 GLS GLS GLS 112...
Страница 113: ...4 2 2 4 2 3 113...
Страница 114: ...4 2 4 1 241 689 kPA 35 100 psi 2 3 1 106 4 5 4 3 4 3 1 USB 3 9 84 106 USB USB 4 3 2 4 3 3 5 5 1 4 1 114...
Страница 115: ...4 1 5 2 6 3 1 7 4 8 1 USB 1 115...
Страница 116: ...1 SST500 SDBS SST8000 6 6 1 1 2 admin 3 password 4 5 6 117 116...
Страница 117: ...7 7 1 7 1 1 USB USB 1 2 2 5 10 30 60 USB 3 7 1 2 1 2 1 3 5 10 30 60 1 15 20 30 45 60 90 120 1 2 3 4 5 5 10 30 60 117...
Страница 118: ...3 7 1 3 120 1 2 3 7 1 4 1 2 1 64 1 5 ID ID 118...
Страница 119: ...4 ppm 2 mL 5 ppm ppb mg L g L psi mbar TIC TIC ppb ppb 3 7 1 5 1 2 AS0640 7 2 7 2 1 1 2 120 USB 119...
Страница 120: ...ID 3 4 7 2 2 1 2 1 5 ID ID ID 3 4 5 6 7 7 3 7 3 1 1 2 PDF CSV XML USB 119 USB USB 120...
Страница 121: ...7 4 MSDS SDS 1 2 SST500 SDBS SST8000 3 Lot ID 2 4 8 MSDS SDS 8 1 1 2 8 2 2 121...
Страница 123: ...5 GLS 6 7 9 UV UV UV UV UV 0 TOC TOC USB USB USB USB USB USB USB 123...
Страница 125: ...9459400 KHP 5 ppm C 9459500 SDBS 9459600 USP 500 ppb 9459700 USP 8 ppm 9459800 9459900 9460000 9460100 125...
Страница 126: ......
Страница 127: ......