Transmitter 9135 - pH measurement in pure water
Page 36
Temperature calibration
To reach the temperature calibration screen, press the
EXECUTION mode with ENTER.
Once the measurement is stable, press
OK
.
You can change the temperature value since the
screen.
Display of motion.
Note :
The temperature adjustment limits are between
- 50°C et + 20°C.
TEMP. CALIB.
EXECUTION
Δ
T : - 2,5°C
Press
ENTER
TEMP. CALIB.
6.8
°C
4.2 pH
TEMP. CALIB.
6.8
°C
06.8°C
TEMP. CALIB.