2—23
User-Entered Calibration,
continued
5.
To edit the S0 value, press
READ/ENTER
. Use the
ZERO/SCROLL
key to select the
numerical value, then press the
READ/ENTER
key to accept the entry and
advance to the next decimal place. Repeat this sequence until the S0
concentration value is entered.
6.
After editing the S0 value, press
READ/ENTER
to accept. “A0” will appear on
the display.
7.
To edit the absorbance value for S0, press the
READ/ENTER
key to go to entry
mode. Use the
ZERO/SCROLL
key to select the numerical value, then press the
READ/ENTER
key to accept the entry and advance to the next decimal place.
Repeat this sequence until the absorbance value for S0 is entered.
8.
After editing A0, press
READ/ENTER
to accept. “S1” will appear on the display.
9.
Repeat steps
4
through
8
for each standard value and absorbance value pair in
the calibration curve.
10. When all calibration data has been reviewed or edited, “ADD” will appear in
the display.
11. Press
READ/ENTER
to add more calibration points, or press
MENU
twice to
return to the measurement mode.
Содержание POCKET COLORIMETER II
Страница 13: ...1 13 Instrument Cap Cord continued Figure 1 Attaching the Instrument Cap Cord...
Страница 14: ...1 14...
Страница 38: ...1 38...
Страница 73: ...1 73 pH continued 10 Press READ ENTER The display will show followed by results in pH units...
Страница 77: ...2 1 Section 2 Instrument Manual...
Страница 78: ...2 2...
Страница 83: ...2 7 Instrument Operation continued Figure 1 Battery Installation...
Страница 84: ...2 8...
Страница 108: ...2 32...
Страница 110: ...2 34...