2—22
User-Entered Calibration,
continued
10. When all the calibration data has been entered, press
MENU
twice to return to
the measurement mode.
Editing a User-entered or Factory Calibration Curve
1.
Press the
MENU
key and hold it down until the display shows “USER”, followed
by “CAL”. Press
ZERO/SCROLL
until EDIT appears.
2.
Press the
READ/ENTER
key to enter Edit mode. In factory-calibrated meters,
“S0” will appear in the display.
Note: When editing a factory-calibrated meter or range, RES (resolution) cannot be changed.
Note: When RES or S0 appears in the display, press
ZERO/SCROLL
to quickly scroll to the
data to be edited.
3.
In uncalibrated meters or in Abs range, “RES” will appear. Press
ZERO/SCROLL
to review the current resolution. Press
ZERO/SCROLL
again to accept the
displayed resolution. To change the resolution (decimal placement), press
READ/ENTER
. Press
ZERO/SCROLL
to select the new resolution, then press
READ/ENTER
to accept. “S0” will appear on the display.
4.
Press
READ/ENTER
. The current concentration value for S0 will appear on the
display.
Содержание POCKET COLORIMETER II
Страница 13: ...1 13 Instrument Cap Cord continued Figure 1 Attaching the Instrument Cap Cord...
Страница 14: ...1 14...
Страница 38: ...1 38...
Страница 73: ...1 73 pH continued 10 Press READ ENTER The display will show followed by results in pH units...
Страница 77: ...2 1 Section 2 Instrument Manual...
Страница 78: ...2 2...
Страница 83: ...2 7 Instrument Operation continued Figure 1 Battery Installation...
Страница 84: ...2 8...
Страница 108: ...2 32...
Страница 110: ...2 34...