10
English
Installation
WARNING
The metering hoses must always be disconnected from the relevant canisters
before any work is carried out to the metering or valve hoses. (Risk of reflux)
Metering hoses are hoses that connect to the sample and reagent pumps. Each metering
hose must only be connected to the applicable sample or reagent pump.
Important note:
Replace the sample metering hose, the discharge hose and the hose for
reagent B after
3 months
. Brush the pump rollers on the sample pump with silicon paste
). Replace the metering hoses for the reagents A, B, C and D after
6 months
.
Valve hoses are connections between the hose pinch valves.
Important note:
Valve hoses must be replaced after
6 months
.
Figure 3 Application of silicon paste
Color code
Designation
Course in instrument
Purple
Discharge hose
Runs from the cuvette intake (
Figure 2
, position
1
, page
9
) via the discharge pump to the exterior
wall
Red
Sample metering hose
Runs from hose pinch valves Q1 and Q2 (T piece
valve hose) via the sample pump to the cuvette
intake (
Figure 2
, position
3
)
Red
Reagent B metering hose
Runs from reagent B canister via the hose pinch
valve Q3 and via the sample pump to the cuvette
intake (
Figure 2
, position
7
)
Yellow
Reagent C metering hose
Runs from reagent C canister via the reagent
pump C, D (dual-channel pump) to the cuvette
intake (
Black
Reagent D metering hose
Runs from reagent D canister via the reagent
pump C, D (dual-channel pump) to the cuvette
intake (
Figure 2
, position
5
)
Orange
Reagent A metering hose
Runs from reagent A canister via the hose pinch
valve QA in the refrigerator and via the reagent A
pump to the cuvette
intake (
, position
). The small air hose at
the bottom end (also marked orange) is inserted
into the front slot.
Red
Valve hose
Feed hose for sample, reagent B, standard
solution and air (
Figure 2
, position
2
).