3.3 Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.

Electrical equipment marked with this symbol may not be disposed of in European
domestic or public disposal systems. Return old or end-of-life equipment to the
manufacturer for disposal at no charge to the user.

3.4 Product hazards

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.

C A U T I O N

 

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.

C A U T I O N

 

Personal injury hazard. Glass components can break. Handle with care to prevent cuts.

Section 4 Preparation for use

New probes come with a soaker bottle that contains storage solution to keep the glass bulb and
reference junction hydrated. Prepare the probe as follows.

1.

Loosen the cap on the soaker bottle. Remove the soaker bottle from the probe.

2.

Rinse the reference junction and glass bulb with deionized water. Blot dry with a lint-free cloth.

3.

For faster stabilization, soak the probe for 3 or more minutes in the sample before use.

4.

Make sure that the meter has the correct date and time settings. The service-life time stamp in
the probe comes from the date and time settings in the meter.

Note: Some meters automatically open the date and time settings when the meter starts for the first time, or
after battery replacement.

5.

Connect the probe to the meter.

Section 5 Calibration

The procedure that follows is applicable to meters that can connect to Intellical pH probes. Refer to
the applicable meter documentation for meter operation and probe-specific settings.

5.1 Calibration notes

Read the notes that follow before calibration:

• Use prepared pH buffer solutions or mix pH buffer powder pillows with deionized water for

calibration. Discard the prepared buffer solutions after each calibration.

English

   

5

Содержание PHC201

Страница 1: ...DOC022 53 80197 PHC201 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ction 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 Section 7 Verify the calibrat...

Страница 4: ...Table of Contents 2...

Страница 5: ...very alkaline samples Figure 1 Probe overview 1 Probe soaker bottle with storage solution 4 Reference junction 2 Probe body 5 Glass bulb and temperature sensor 3 Probe soaker bottle holder Section 2...

Страница 6: ...ufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water qua...

Страница 7: ...e with a soaker bottle that contains storage solution to keep the glass bulb and reference junction hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...ads each data point and when the meter is set to read at regular intervals The user must manually save each data point when the meter is set to read continuously Air bubbles below the sensor can cause...

Страница 10: ...re 1 Go to the verification menu The display shows the pH buffer solution to use for verification Note Menu name for HQd meters Run check standard 2 Prepare or pour the pH buffer solution into a beake...

Страница 11: ...body with a lint free cloth Table 1 Cleaning solution Contamination Cleaning solution Active component Soak time General contamination Electrode cleaning solution for regular maintenance KATHON CG DEC...

Страница 12: ...reference junction below the liquid level in the soaker bottle Add storage solution to the soaker bottle if necessary Figure 2 Probe storage Section 9 Troubleshooting Keep the probe clean and in the r...

Страница 13: ...erence junction is not sufficient Shake the probe in the solution from side to side to refresh the reference junction The stir speed is too slow or too fast Try a different stir speed An incorrect buf...

Страница 14: ...r powder pillow NIST 50 pkg 2226966 pH 7 00 0 02 pH buffer powder pillow NIST 50 pkg 2227066 pH 10 01 0 02 pH buffer powder pillow NIST 50 pkg 2227166 Radiometer Analytical IUPAC Series certified pH s...

Страница 15: ...older 30 mL 1 923 656 Beaker holder 30 mL 1 923 556 Beaker 100 mL polypropylene 1 108042 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard In...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: