Section 8 Maintenance

Regular maintenance is necessary for the best accuracy, stabilization time and life of the probe.
Keep the probe in the recommended storage solution between measurements.

8.1 Clean the probe

Clean the probe regularly to remove contamination and to keep the reference junction open.
Symptoms of contamination:

• Incorrect or irregular readings
• Slow stabilization times
• Calibration errors
• Sample material stays on the probe

1.

Rinse the probe with deionized water. Use warm (35–45 °C (95–113 °F)) deionized water to
remove storage solution that dries on the probe. Dry the probe body with a lint-free cloth.

2.

Soak the glass bulb and reference junction in the applicable cleaning solution for the specified
time. Refer to 

Table 1

 and 

Consumables

 on page 11.

3.

Rinse or soak the probe for 1 minute in deionized water. Dry the probe body with a lint-free cloth.

4.

Soak the probe in pH 4 buffer for 20 minutes.

5.

Rinse the probe with deionized water. Dry the probe body with a lint-free cloth.

Table 1  Cleaning solution

Contamination

Cleaning solution

Active component

Soak time

General
contamination

Electrode cleaning solution
for regular maintenance

KATHON

 CG,

DECONEX

®

11 

12–16 hours

Minerals

Electrode cleaning solution
for minerals/inorganic
contamination

Phosphoric acid
(~10%)

10–15 minutes

Fats, grease and oils

Electrode cleaning solution
for fats, oils and grease
contamination

KATHON

 CG,

TRITON

®

 X

2 hours maximum

Proteins

Electrode cleaning solution
for proteins/organic
contamination

Pepsin in HCl

3 hours maximum

Wastewater and
organic compounds

Electrode cleaning solution,
extra strong

Sodium hypochlorite

5–10 minutes

Blockage of porous
pin/diaphragm
reference junctions

Electrode cleaning solution
for porous pin/diaphragm
reference junctions

Thiourea

3 hours with stirring
or until the porous
plug becomes white

8.2 Soak procedure for dry probes

If the glass bulb becomes dry, complete the steps that follow to hydrate the probe.

1.

Soak the probe tip in pH 4 and pH 7 buffer solutions for 5 minutes in each solution.

2.

Rinse the probe with deionized water. Blot dry with a lint-free cloth.

3.

Calibrate the probe.

English

   

9

Содержание PHC201

Страница 1: ...DOC022 53 80197 PHC201 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ction 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 Section 7 Verify the calibrat...

Страница 4: ...Table of Contents 2...

Страница 5: ...very alkaline samples Figure 1 Probe overview 1 Probe soaker bottle with storage solution 4 Reference junction 2 Probe body 5 Glass bulb and temperature sensor 3 Probe soaker bottle holder Section 2...

Страница 6: ...ufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water qua...

Страница 7: ...e with a soaker bottle that contains storage solution to keep the glass bulb and reference junction hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...ads each data point and when the meter is set to read at regular intervals The user must manually save each data point when the meter is set to read continuously Air bubbles below the sensor can cause...

Страница 10: ...re 1 Go to the verification menu The display shows the pH buffer solution to use for verification Note Menu name for HQd meters Run check standard 2 Prepare or pour the pH buffer solution into a beake...

Страница 11: ...body with a lint free cloth Table 1 Cleaning solution Contamination Cleaning solution Active component Soak time General contamination Electrode cleaning solution for regular maintenance KATHON CG DEC...

Страница 12: ...reference junction below the liquid level in the soaker bottle Add storage solution to the soaker bottle if necessary Figure 2 Probe storage Section 9 Troubleshooting Keep the probe clean and in the r...

Страница 13: ...erence junction is not sufficient Shake the probe in the solution from side to side to refresh the reference junction The stir speed is too slow or too fast Try a different stir speed An incorrect buf...

Страница 14: ...r powder pillow NIST 50 pkg 2226966 pH 7 00 0 02 pH buffer powder pillow NIST 50 pkg 2227066 pH 10 01 0 02 pH buffer powder pillow NIST 50 pkg 2227166 Radiometer Analytical IUPAC Series certified pH s...

Страница 15: ...older 30 mL 1 923 656 Beaker holder 30 mL 1 923 556 Beaker 100 mL polypropylene 1 108042 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard In...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...

Отзывы: