Hach PHC201 Скачать руководство пользователя страница 1

DOC022.53.80197

PHC201

05/2021, Edition 3

User Manual

Содержание PHC201

Страница 1: ...DOC022 53 80197 PHC201 05 2021 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...ction 5 Calibration 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 Section 7 Verify the calibrat...

Страница 4: ...Table of Contents 2...

Страница 5: ...very alkaline samples Figure 1 Probe overview 1 Probe soaker bottle with storage solution 4 Reference junction 2 Probe body 5 Glass bulb and temperature sensor 3 Probe soaker bottle holder Section 2...

Страница 6: ...ufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water qua...

Страница 7: ...e with a soaker bottle that contains storage solution to keep the glass bulb and reference junction hydrated Prepare the probe as follows 1 Loosen the cap on the soaker bottle Remove the soaker bottle...

Страница 8: ...remove air bubbles during calibration The pH buffer solutions have known pH values at different temperatures The meter uses the mV and temperature readings of the probe in the pH buffer solutions to...

Страница 9: ...ads each data point and when the meter is set to read at regular intervals The user must manually save each data point when the meter is set to read continuously Air bubbles below the sensor can cause...

Страница 10: ...re 1 Go to the verification menu The display shows the pH buffer solution to use for verification Note Menu name for HQd meters Run check standard 2 Prepare or pour the pH buffer solution into a beake...

Страница 11: ...body with a lint free cloth Table 1 Cleaning solution Contamination Cleaning solution Active component Soak time General contamination Electrode cleaning solution for regular maintenance KATHON CG DEC...

Страница 12: ...reference junction below the liquid level in the soaker bottle Add storage solution to the soaker bottle if necessary Figure 2 Probe storage Section 9 Troubleshooting Keep the probe clean and in the r...

Страница 13: ...erence junction is not sufficient Shake the probe in the solution from side to side to refresh the reference junction The stir speed is too slow or too fast Try a different stir speed An incorrect buf...

Страница 14: ...r powder pillow NIST 50 pkg 2226966 pH 7 00 0 02 pH buffer powder pillow NIST 50 pkg 2227066 pH 10 01 0 02 pH buffer powder pillow NIST 50 pkg 2227166 Radiometer Analytical IUPAC Series certified pH s...

Страница 15: ...older 30 mL 1 923 656 Beaker holder 30 mL 1 923 556 Beaker 100 mL polypropylene 1 108042 Disposable wipes 11 x 22 cm 280 pkg 2097000 Wash bottle polyethylene 500 mL 1 62011 Probe stand for standard In...

Страница 16: ...14 English...

Страница 17: ......

Страница 18: ...2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 ro...