33
Installation
3.9 Sensor installation
3.9.1 Sensor positioning information
The sensor must be installed in a socket or flow chamber that allows contact with the sample
fluid to be analyzed.
The sensor and measuring instrument are connected by a cable. The standard sensor cable
lengths are 3, 5, 10, 15 and 20 meters.
Ensure that the sensor will be mounted:
•
perpendicular to the pipe
•
horizontal
•
on a horizontal pipe section (or on flow-ascending vertical pipe)
•
minimum of 15 meters away from the pump's discharge side
•
in a place where the sample flow is stable and rapid, and as far as possible from:
•
valves
•
pipe bends
•
the suction side of any pumps
•
a CO
2
injection system or similar
Note:
There may be situations where not all the above conditions can be met. If this is the case, or you
have any concerns, please consult your Hach representative to appraise the situation and define the best
applicable solution.
3.9.2 Recommended sample flow rate
For optimal response time, the recommended sample flow rate for both the K1100 and M1100
sensors is 150 mL/min.
Control the flow with the outlet valve on the flow meter (number 3 in
below) to avoid
foam developing inside the flow chamber which could lead to erroneous oxygen measurement
values.
3.9.3 Sensor insertion
•
Insert the sensor straight into the flow chamber or socket. Do not twist the sensor.
•
Hand tighten the attaching collar.
•
Connect the sensor cable.
•
Check for leaks; replace O-rings if product leaks are visible.
Figure 19 Typical measurement schematic
1.
Valve to redirect sample for measurement
3.
Flow meter with outlet valve
2.
Sensor and flow chamber combination
4.
Drain
Содержание ORBISPHERE K-M1100
Страница 1: ...DOC024 52 93023 ORBISPHERE Model K M1100 Sensor and Model 410 Analyzer USER MANUAL 01 2021 Edition 18 ...
Страница 5: ...4 Table of Contents ...
Страница 19: ...18 Specifications ...
Страница 39: ...38 Installation ...
Страница 45: ...44 User Interface ...
Страница 48: ...47 Section 6 View Menu Figure 33 View menu ...
Страница 51: ...50 View Menu ...
Страница 52: ...51 Section 7 Measurement Menu Figure 36 Measurement menu ...
Страница 59: ...58 Measurement Menu ...
Страница 65: ...64 Calibration Menu ...
Страница 66: ...65 Section 9 Inputs Outputs Menu Figure 39 Inputs Outputs menu ...
Страница 68: ...67 Inputs Outputs Menu 9 4 Analog outputs Figure 40 Analog outputs menu ...
Страница 75: ...74 Inputs Outputs Menu ...
Страница 91: ...90 Communication Menu ...
Страница 94: ...93 Section 12 Products Menu Figure 43 Products menu ...
Страница 97: ...96 Global Configuration Menu ...
Страница 98: ...97 Section 14 Services Menu Figure 45 Services menu part 1 ...
Страница 99: ...98 Services Menu Figure 46 Services menu part 2 ...
Страница 107: ...106 Maintenance and Troubleshooting ...
Страница 113: ...112 Accessories and Spare Parts ...