45
Measurement Menu
Measurement alarms configuration
Set the thresholds for the low/high concentration levels, according to the application. Each alarm type
can be individually enabled or disabled without losing its settings.
These events can activate the relays and can be displayed.
•
Low-low
: 2nd stage for too low concentration
•
Low
: 1st stage for too low concentration
•
High
: 1st stage for too high concentration
•
High-high
: 2nd stage for too high concentration
•
Hysteresis
: A percentage of the above concentration values. The hysteresis is used to prevent relay
“flickering” when the measurement is just at the alarm levels. Set this to a minimum, but enough to
eliminate flickering.
As an example, if the High Alarm is set to 40 units and the Hysteresis is set to 10%, then the High
Alarm is activated once the measurement reaches 40 units, but only deactivated once the
measurement drops below 36 units. With the Low Alarm the opposite is true, in that if the Low Alarm
is set to 20 units and the Hysteresis set to 10%, then the Low Alarm is activated when the
measurement drops below 20 units, and deactivated when the measurement rises above 22 units.
•
Delay
: The delay in seconds, before alarms go on whenever concentration values go above ”High
alarms” or below “Low alarms”. Set this to a minimum value, but enough to avoid alarms for
non-representative peaks beyond the set level.
Figure 29 Alarms system diagram
Measurement filter configuration
The filters are aimed at “flattening” the measurement curve in situations where the process shows
atypical peak values that could otherwise hamper the interpretation of measurement readings. The filter
is applied on the last set of measurements each time a measurement is taken.
•
Mean
: Mathematical average of the last set (depth) of measurement values.
•
Median filter
: Allows for eliminating atypical peak measurement values, and average the remaining
ones. The calculation sorts the last measurements set (depth) by values, then delete the highest and
lowest values, and averages the remaining values (central depth).
Example for depth 7, central depth 5:
Sorted values, both ends eliminated, the average of the center five is then
3.88
.
0.7
- 1.1 - 4.0 - 4.3 - 4.4 - 5.6 -
7.0
Example for depth 5, central depth 3:
Sorted values, both ends eliminated, the average of the center three is then
4.23
.
1.1
- 4.0 - 4.3 - 4.4 -
5.6
Example for depth 8, central depth 4:
Sorted values, both ends eliminated, the average of the center four is then
4.43
.
0.7 - 1.1
- 4.0 - 4.3 - 4.4 - 5.0 -
5.6 - 7.0
Содержание ORBISPHERE 410
Страница 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Страница 5: ...4 Table of Contents...
Страница 13: ...12 General Information...
Страница 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Страница 39: ...38 User Interface...
Страница 40: ...39 Section 5 View Menu Figure 25 View menu...
Страница 43: ...42 View Menu...
Страница 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Страница 49: ...48 Measurement Menu...
Страница 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Страница 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Страница 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Страница 65: ...64 Inputs Outputs Menu...
Страница 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Страница 85: ...84 Global Configuration Menu...
Страница 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Страница 87: ...86 Services Menu Figure 39 Services menu Part 2...
Страница 91: ...90 Services Menu...
Страница 95: ...94 Part Lists...