17
Section 3
Installation
This section provides necessary information to install and connect the analyzer. The installation
of the analyzer should be performed in accordance with relevant local regulations.
3.1 Unpacking
Remove carefully the instrument and its accessories from the box and packing material,
referring to the packing list included to confirm that everything has been delivered.
Please visually inspect the instrument for shipping damage. If anything is missing or damaged,
contact the manufacturer or your dealer immediately.
You may want to retain the box and other packing material in case later you need to ship the
instrument. Refer to
Storage, handling and transportation on page 92
. Please dispose safely
and ecologically of the box and packing material (if not stored for future use).
Please read through this manual thoroughly before carrying out the installation.
DANGER
Electrocution Hazard. Do not connect AC power to a 5 VDC powered model.
WARNING
Potential Electrocution Hazard. Always disconnect power to the instrument when
making electrical connections.
WARNING
Potential Electrocution Hazard. If this equipment is used outdoors or in potentially
wet locations a
Ground Fault Interrupt
device must be used for connecting the
equipment to its main power source.
WARNING
Potential Electrocution Hazard. A protective earth (PE) ground connection is
required for both 100-240 VAC and 5 VDC wiring applications. Failure to connect a
good PE ground connection can result in shock hazards and poor performance due
to electromagnetic interferences. ALWAYS connect a good PE ground to the
controller terminal.
CAUTION
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this
section of the manual.
NOTICE
Install the device in a location and position that gives easy access to the disconnect device
and its operation.
NOTICE
Potential Instrument Damage. Delicate internal electronic components can be
damaged by static electricity, resulting in degraded performance or eventual failure.
Содержание ORBISPHERE 410
Страница 1: ...DOC024 52 93001 ORBISPHERE Model 410 Analyzer USER MANUAL 11 2017 Edition 15...
Страница 5: ...4 Table of Contents...
Страница 13: ...12 General Information...
Страница 23: ...22 Installation Figure 5 Panel mount instrument dimensions in millimeters...
Страница 39: ...38 User Interface...
Страница 40: ...39 Section 5 View Menu Figure 25 View menu...
Страница 43: ...42 View Menu...
Страница 44: ...43 Section 6 Measurement Menu Figure 28 Measurement menu...
Страница 49: ...48 Measurement Menu...
Страница 51: ...50 Calibration Menu Figure 31 Calibration menu cont from previous page...
Страница 56: ...55 Section 8 Inputs Outputs Menu Figure 32 Inputs Outputs menu...
Страница 58: ...57 Inputs Outputs Menu 8 4 Analog outputs Figure 33 Analog outputs menu...
Страница 65: ...64 Inputs Outputs Menu...
Страница 82: ...81 Section 11 Products Menu Figure 36 Products menu...
Страница 85: ...84 Global Configuration Menu...
Страница 86: ...85 Section 13 Services Menu Figure 38 Services menu Part 1...
Страница 87: ...86 Services Menu Figure 39 Services menu Part 2...
Страница 91: ...90 Services Menu...
Страница 95: ...94 Part Lists...