Attach compressed air (not for flow cell operation)
N O T I C E
Make sure that the air pressure is not more than 7 bar or damage to the valve can occur.
The probe has measuring windows with nano-coating and a compressed air cleaning system. Both
components are important to prevent particles in the optical path or fouling of the measuring
windows.
1.
Use a 13 mm (0.5 in.) wrench to install the compressed air connector on the probe. Refer to
2.
Attach the compressed air tube to the compressed air connector. Refer to
.
3.
For the compressed air line, install the High Output Airblast System (HOAB), the optional High
Pressure Air Cleaning System or a magnetic valve/external compressed air supply. Set the air
pressure between 3 and 6 bar. Make sure that the total length of the compressed air line is less
than 25 m (82 ft).
Note: Do not make a measurement during the cleaning interval. When compressed air supply is controlled by
the controller, measurement and cleaning are synchronized. Refer to
Set the automatic cleaning options
on page 13.
Note: Add a check valve in the compressed air line to prevent clogging in the compressed air connector.
Figure 8 Compressed air connection
1
Compressed air connector
3
Optical windows
2
Compressed air line
4
Optical path
12
English
Содержание NV3300
Страница 2: ...English 3 Fran ais 19 Espa ol 35 51 67 Portugu s 82 98 114 2...
Страница 53: ...ESD 53...
Страница 54: ...NV3300 SAC Spectral Absorption Coefficient 1 8 LED 4 50 100 150 250 mm 51 CD500 CD300 54...
Страница 55: ...1 1 4 2 5 3 6 2 2 1 2 55...
Страница 57: ...OFF 3 4 5 3 57...
Страница 58: ...4 1 6 mm 3 2 4 8 mm 5 58...
Страница 59: ...M12 10 m 6 7 COM COM 6 7 3 1 RS232 RX RS485 A 2 3 ETH_TX 4 ETH_TX 5 12 24 VDC 6 ETH_RX 59...
Страница 64: ...MSDS SDS 1 Options 2 Service mode HOLD 3 4 5 Service mode HOLD 1 1 2 2 1 62 2 3 4 64 5 64...
Страница 69: ...ESD TV 69...
Страница 70: ...NV3300 SAC 1 8 LED 4 50 100 150 250 mm 67 CD500 CD300 1 1 4 2 5 3 6 2 70...
Страница 72: ...2 SAC 4 LED 1 nm LED 2 nm SAC 436 1 m 436 740 DIN EN ISO 7887 2011_method B NBR OFF 3 4 5 72...
Страница 73: ...3 4 1 6 mm OD 3 2 4 8 mm OD 73...
Страница 74: ...5 M12 10 m 6 7 COM COM 6 74...
Страница 76: ...8 1 3 2 4 1 Options 2 Cleaning 3 15 5 10 mA 5 Execute Sensor 5 76...
Страница 79: ...MSDS SDS 1 Options 2 Service mode HOLD 3 4 5 Service mode HOLD 1 2 1 77 2 3 4 79 5 79...
Страница 100: ...ESD 100...
Страница 101: ...NV3300 SAC 1 8 LED 4 50 100 150 250 98 CD500 CD300 101...
Страница 102: ...1 1 4 2 5 3 6 2 2 1 2 102...
Страница 104: ...OFF 3 4 5 3 104...
Страница 105: ...4 1 OD 6 3 2 4 OD 8 5 105...
Страница 106: ...10 M12 6 7 COM COM 6 7 M12 3 8 1 RS232 RX RS485 A 2 3 ETH_TX 4 ETH_TX 5 Power 12 24 VDC 6 ETH_RX 106...
Страница 107: ...3 8 7 ETH_RX 8 RS232 TX RS485 B 7 1 13 0 5 8 2 8 3 HOAB 3 6 25 82 108 8 1 3 2 4 107...
Страница 111: ...1 Options 2 Service mode HOLD 3 4 5 Service mode HOLD 1 1 2 2 1 109 2 3 4 110 5 111...
Страница 116: ...ESD 116...
Страница 117: ...NV3300 SAC 1 8 LED 4 50 100 150 250 mm 114 CD500 CD300 1 1 4 2 5 3 6 2 117...
Страница 119: ...NBR OFF 3 4 5 3 119...
Страница 120: ...4 1 6 mm 3 2 4 8 mm 5 120...
Страница 121: ...10 M12 6 7 COM COM 6 7 M12 3 8 1 RS232 RX RS485 A 2 3 ETH_TX 4 ETH_TX 5 12 24 V 6 ETH_RX 121...
Страница 122: ...3 8 7 ETH_RX 8 RS232 TX RS485 B 7 bar 1 13 mm 0 5 in 8 2 8 3 HOAB 3 6 bar 25 m 82 ft 123 8 1 3 2 4 122...
Страница 126: ...3 4 5 Service mode HOLD 1 1 2 2 1 124 2 3 4 125 5 Range 1 125 2 114 3 124 4 5 6 1 2 3 NV3300 123 4 5 90 90 126...
Страница 128: ......
Страница 129: ......