Troubleshooting
6.3.2 Troubleshooting during calibration
Symptom
Possible cause
Corrective actions
SENSOR CODE
Sensor code entered incorrectly
Using the certificate, check whether the
sensor code was entered correctly. If not
having the certificate, carry out "Factory
Calibration" (Factory cal).
NITRATE
OFFSET
Error in the last nitrate calibration, sensor
cartridge too old, contaminated, faulty
Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.
SLOPE
Error in the last nitrate calibration, sensor
cartridge too old, contaminated, faulty
Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.
CHLORIDE
OFFSET
Error in the last chloride calibration, sensor
cartridge too old, contaminated, faulty
Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.
SLOPE
Error in the last chloride calibration, sensor
cartridge too old, contaminated, faulty
Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.
Содержание NO3D sc
Страница 2: ...2...
Страница 4: ...4 Table of Contents Section 10 Certification 43 Appendix A Modbus Register 45 Index 47...
Страница 6: ...6 Specifications 1 1 Dimensions Figure 1 Stainless steel sensor dimensions...
Страница 10: ...10 General Information...
Страница 16: ...16 Installation Figure 10 Example of sensor installation with Rail Mount Kit...
Страница 18: ...18 Installation...
Страница 28: ...28 Operation...
Страница 38: ...38 Spare Parts and Accessories...
Страница 44: ...44 Certification...
Страница 48: ...48...