Option
Description
Server Port
ModbusTCP server listen port (default = 502). Range:
0 to 65535, integer only.
Client Port
Not available (disabled)
Gateway
Router or access point to another network (default =
169.254.1.5)
Remote Server
IP
Not available (disabled)
Ethernet
Protocol
Sets the Ethernet protocol to Modbus or FXB.
2.
For Ethernet units with the analog output feature, refer to
the LAN settings through a network
on page 19 to configure the
Ethernet settings.
3.
Click
Save Settings
to save the changes.
Configure the LAN settings through a network
1.
In the setup utility software, select the
LAN Setup
tab. The software
looks for LAN instruments. The LAN instruments found are shown.
2.
Select an instrument to show the LAN instrument settings.
3.
Change the LAN settings. Refer to the options table in
on page 18.
4.
Click
Save Settings
to save the changes.
Do an analog output test
For instruments with the analog output feature, do an analog output test.
1.
Connect the analog outputs to the load resistors of the data
acquisition system.
Note: As an alternative, install a set of load resistors with 0.1% accuracy and
at least 0.25 W capability across the analog output. Load resistor values of
100, 250 or 500 ohms are typically used.
2.
Let a tiny amount of particles flow through the instrument to get a
count in the test channel.
Note: One method to get counts is to use a zero count filter, and put a pin-hole
in the tubing that is between the filter and the instrument.
3.
On the Basic Setup tab of the setup utility software, temporarily set:
• Count Cycles—1
• Sample Timing: Hold—10 seconds or more
4.
Click
Save Settings
.
5.
Select the Data Display tab, then click
Monitor
if shown so the data
shown can update continuously as each sample is taken.
6.
Click
Sample
if shown to start sample collection.
7.
When the Status value changes from "Count" to "Stop", measure the
voltage across the load resistors for each channel. Also note the
counts shown in the display for each channel.
8.
Use the equation that follows to calculate the expected voltage from
the counts shown. Make sure that the measured and calculated
voltages agree.
Voltage = (((SC ÷ FC × 16) + 4) ÷ 1000) × RL
Where:
SC = sample count at the end of the sample period
FC = full-scale channel count. Refer to the analog settings in the
setup utility software.
RL = value of the load resistor in ohms
The expected output voltage when the full-scale channel count is
1000 with a 100, 250 and 500-ohm resistor is shown in
.
9.
To do the test again, do steps
–
.
10.
For units with a flow monitor, temporarily remove the central vacuum
from the instrument.
11.
While a flow alarm is active, measure the voltage across the load
resistors for each channel.
12.
Use the equation that follows to calculate the expected voltage.
Make sure that the measured and calculated voltages agree.
Voltage = < (0.002 × RL)
Where: RL = value of the load resistor in ohms
Example: for a 100-ohm resistor, the voltage should be less than
0.20 V.
English
19
Содержание MET ONE 6003
Страница 2: ...English 3 Fran ais 24 Espa ol 48 72 91 113 2...
Страница 13: ...Figure 9 5 pin connector wiring Figure 10 10 pin connector wiring English 13...
Страница 35: ...Figure 9 C blage du connecteur 5 broches Figure 10 C blage du connecteur 10 broches Fran ais 35...
Страница 59: ...Figura 9 Cableado del conector de 5 patillas Figura 10 Cableado del conector de 10 patillas Espa ol 59...
Страница 73: ...5 1 CE 1 ISO21501 4 2005 8 12 2002 96 EC 1 CDRH 9022243 029 2007 6 24 50 IEC EN 60825 1 21 CFR 1040 10 73...
Страница 75: ...1 3 1 1 ModBus FX 2 Status 10Base T 100Base T 2 2 1 3 2 4 3 3 75...
Страница 77: ...3 m 10 ft 3 m 10 ft 1 m 3 m 10 ft schedule 80 PVC Cobolite DIN DIN 5 77...
Страница 78: ...5 DIN 79 6 90 90 T 78...
Страница 81: ...9 5 10 10 81...
Страница 83: ...12 1 5 2 6 4 20 mA 3 24 VDC 7 24 VDC 4 8 13 1 3 2 4 VHP IPA 83...
Страница 89: ...3 PC CSV 15 1 1 7 cfm L min 2 8 cfm L min 3 9 RH 4 10 5 11 6 1 89...
Страница 90: ...9 9 1 90 X 90 X FMS 90...
Страница 92: ...6003 6005 140 000 000 m3 4 000 000 ft3 10 6013 6015 20 000 000 m3 566 000 ft3 10 5 1 CE 1 ISO21501 4 Web CD 92...
Страница 94: ...1 1 10 1 7 RH 2 8 3 9 2 2 4 1 10 RJ45 2 5 11 5 2 6 DIP 1 1 2 1 3 1 1 1 1 FX ModBus 2 10Base T 100Base T 94...
Страница 95: ...2 2 1 3 2 4 3 3 3 1 3 5 2 4 3 0 0 1 cfm 0 3 m 0 5 m F 1 1 0 cfm 0 5 m N 95...
Страница 97: ...3 m 3 m 1 m 3 m 80 PVC Cobolite DIN DIN 5 97...
Страница 98: ...5 DIN 99 6 90 90 T 98...
Страница 99: ...6 1 3 2 90 4 T 2 3 m2 1 2 2 95 30 cm 30 cm 4 4 Hytrel Bevaline Tygon Hytrel Bevaline Tygon 99...
Страница 100: ...1 3 m 2 3 4 1 2 m 10 cm 5 6 24 VDC 5 10 7 5 8 6 28 VDC 5 9 10 10 7 5 5 5 1 4 9 28 VDC 1 A 2 5 3 8 10 100...
Страница 112: ...HACH Japan 9 9 1 112 X 112 X FMS 1 112...
Страница 114: ...5 1 CE 1 ISO21501 4 CD 114...
Страница 116: ...1 1 10 1 7 RH 2 8 3 9 2 2 4 1 10 RJ45 2 5 11 5 2 6 DIP 1 1 2 1 3 1 1 1 1 1 FX ModBus 2 10Base T 100Base T 2 116...
Страница 119: ...DIN DIN 5 5 DIN 119...
Страница 120: ...120 6 90 90 T 6 1 3 2 90 4 T 2 3 m2 25 ft2 1 120...
Страница 123: ...9 5 10 10 123...
Страница 125: ...12 1 5 2 6 4 20 mA 3 24 VDC 7 24 VDC 4 8 13 1 3 2 4 VHP IPA 125...
Страница 133: ...9 9 1 133 X 133 X FMS 1 133...
Страница 134: ......
Страница 135: ......