Operation
5.1.3
Manual calibration value correction
1.
Select
MENU
, confirm with
ENTER/ON
.
2.
Select
CALIBRATE
, confirm with
ENTER/ON
.
The instrument reads the probe data.
3.
Select the curve shown and confirm with
ENTER/ON
.
4.
Select the desired curve and confirm with
ENTER/ON
.
5.
Select the desired point and confirm with
ENTER/ON
.
6.
Overwrite the existing calibration point.
7.
Enter the value for the calibration point using the navigation keys and confirm with
ENTER/ON
.
•
Navigation key
LEFT/RIGHT
: Jump to next/previous decimal place
•
Navigation key
UP/DOWN
: Change number
5.1.4
Delete calibration point
Same procedure as in
section 5.1.3
. Set the concentration of the calibration point to the
value 0.000 and confirm with
ENTER/ON
. The calibration point is deleted.
5.1.5
Reset to default calibration
If the turbidity calibration was changed, it can be reset to the default calibration in
accordance with ISO 7027.
Note:
Stop the current measurement before the default calibration is reset.
1.
Select
MENU
, confirm with
ENTER/ON
.
2.
Select
CALIBRATE
, confirm with
ENTER/ON
.
The instrument reads the probe data.
3.
Select turbidity curve
C-Tu
and confirm with
ENTER/ON
.
Note:
If turbidity curve
C-Tu
is not shown in the menu, select the curve shown and confirm with
ENTER/ON
. The selection menu for the curves is shown. Select curve
C-Tu
and confirm with
ENTER/ON
. Curve C-TU is activated.
Open the selection menu for curve C-TU with
ENTER/ON
, select
RESET
and confirm with
ENTER/ON
.
4.
Press navigation key
UP/DOWN
, select
RESET
and confirm with
ENTER/ON
.
The DS curves can also be reset to the default calibration (1-point calibration at 4 g/L).
However, a precise determination of the solids concentration can only be obtained through
calibration (refer to
section 5.1, page 23
).
5.2
Measure
Note:
In order to obtain precise solids measurement results, you must calibrate the instrument
before the first measurement is taken.
Without any calibration of the DS curves, the instrument returns to 1-point default calibration at
4 g/L. Deviations from the actual measurement value may occur.