Overview
6
Import Lot-Specific Calibration Data with New Cartridge Lots
Each test cartridge lot is factory calibrated. Lot-specific calibration data must be loaded from the
USB flash drive that is included with each box of test cartridges. If the file is missing, you will get
a message to load the calibration data from the USB flash drive. Alternatively, calibration data
can be downloaded from the internet by following the prompts in the LightDeck-T software.
1. Turn on the computer
2. Start LightDeck-T software by double clicking the LightDeck-T icon on the desktop to
launch the software.
3. Import the calibration file
4. Select IMPORT.
5. Navigate to the USB drive. Double click the xxx.cdf.xml file. Alternatively, select
INTERNET to download the file.
6.
Select “Update Now” to import the calibration information.
The LightDeck-T software will import the cartridge definition file and calibration information. Do
not remove the USB drive until the screen
shows “Completed successfully.”