33
Main Menu
5.4 Calibration
5.4.1 Barometric sensor calibration
5.4.2 Gas sensor calibration
There are two calibration modes available
•
Zero calibration
•
High level adjustment
The zero calibration method is the best method to guarantee the sensor specifications.
To get more accurate measurements for samples with higher oxygen concentrations (above 1%
oxygen which corresponds to about 400 ppb dissolved O
2
) a high level adjustment can be
performed using a gas mixture containing 2% oxygen. However, this should not be done without
first ensuring the zero point is accurate (i.e. by performing a zero calibration first).
Zero calibration
High level adjustment
There are three possible modes for high level adjustments:
•
Using a reference gas mixture (gas cylinder)
•
Using a known oxygen sample (reference sample)
•
Enter a factory parameter provided by Hach (this option is
only
required when the sensor
spot has been replaced)
If the calibration mode displayed is not the required mode, then press the
Enter
key and select
the required mode from the three options available in the drop-down list:
•
Factory parameter
•
Reference sample
•
Ref. gas bottle
Barometer Calibration
The upper box shows the current barometric pressure as measured by the instrument. Using a precision
certified barometer, measure the barometric pressure in the location where the instrument is used. If the
values differ, enter the correct value in the box provided and select
Validate calibration
O
2
zero calibration
1.
Rinse flow path with 20 mL of ethanol (EtOH) using the syringe and connectors supplied in the tool
kit
2.
Connect a cylinder of oxygen free gas (minimum 99.999% purity) to the instrument and adjust the
flow to approximately 100 mL/min
3.
Let the oxygen free gas run through the instrument for 5 minutes
4.
Press the
Enter
key to start the calibration
5.
Wait until the
Signal in range
and
Stability reached
fields display
Yes
which indicates the
calibration is within acceptable limits. The
Calibration possible
field should also display
Yes
at this
point. Press the
Enter
key to complete the calibration
6.
Accept (
OK
) or reject (
Cancel
) the new calibration data
O2 high level adjustment - Factory parameter
1.
In the
Calibration mode
window select
Factory parameter
2.
Scroll down to the
Reference value
window and enter the factory parameter as found on the
package of the LDO spot (model number DG33218)
3.
Scroll down to the
Start Calibration
window and press the
Enter
key to start the calibration
4.
Accept (
OK
) or reject (
Cancel
) the new calibration data