Page 14
Plumbing Connections
54400_install.fm
Section 1
Figure 3
CL17 Instrument Dimensions (3 of 3)
1.4
Plumbing Connections
Note:
The sample drain contains
analysis wastes, which include both
sample and chemical reagents. The
enclosure drain produces no waste
under normal use. However, in the
event of spills or leaks, some waste
may be produced. Although the
chemicals used are of low
concentration, it is best to check
with local authorities concerning
proper disposal of waste from this
product.
Sample inlet and drain connections are made on the bottom of the instrument
using a quick connect fitting for ¼-inch OD tubing. See
. Connect the
¼-inch O.D. poly tubing by pushing it into the fitting. Two distinct “stops” will be
felt as the tubing is correctly attached. If the tubing is not properly seated, it
will come loose when water pressure is applied. The sample drain fitting is
sized for ½-inch ID flexible tubing (not supplied).
Sample Inlet Quick Connect
.25 O.D. Polyethylene Tubing
(Cat. No. 47438-00)
.50 I.D. Drain Tube Sample
Overflow Drain
Recommended
.50 I.D. Tubing
15.21 (386.30 mm)
″
0.06 (1.59 mm)
Clearance
For Vents
″
7.06
(179.39 mm)
″
19.95
(506.78 mm)
″
R 14.89
(378.18 mm)
″
Drawing Number: 54400-87
Revision: A, 02-16-00
!
Содержание Lange CL17
Страница 1: ...Catalog Number 54400 18 CL17 CHLORINE ANALYZER INSTRUMENT MANUAL 06 04 4ed...
Страница 2: ...Hach Company 2001 2004 All rights reserved Printed in the U S A...
Страница 10: ...Visit http www hach com...
Страница 12: ...Visit http www hach com...
Страница 32: ...Visit http www hach com...
Страница 46: ...Visit http www hach com...
Страница 54: ...Visit http www hach com...
Страница 58: ...Visit http www hach com...
Страница 62: ...Visit http www hach com...
Страница 68: ...Visit http www hach com...