52
SECTION 5,
continued
5.2.1.1
Assembling the Support Stand
1. Verify the sample compartment is empty, and turn the instrument off.
2. Turn the instrument on its top (place it on a soft cloth to protect the instrument
from marring), and install the base plate of the stand as illustrated in Figure 10.
DO NOT OVERTIGHTEN THE SCREWS
.
3. Place the instrument right side up.
4. Install the Flow-Cell Inlet Reservoir on to the support rod.
5. Slide the Collection Drain Assembly on to the support rod.
6. Place the instrument right-side up, and install the support rod into the base.
Tighten the thumb screw to secure the rod.
Figure 10
Base Plate Installation
Base