If this is the first time you have run this test on your DR/2010,
you must contact your local 

Hach

distributor for an upgrade

of your software or perform the Manual Program Update for
the DR/2010 (see Instrument Methods).

INSTRUMENT SET-UP

1. The display should read: 

Enter Program #

2. Enter 

276

on the keypad.

3. Rotate the wavelength dial until the small display shows:

485 nm

.

When the correct wavelength is dialed in, the display will
quickly show: 

Zero Sample

.

4. Insert the COD Vial Adapter into the cell holder with the

marker to the right.

5. Follow the procedure steps.

(

Note:

If measuring Total Iron and performed the sample 

digestion using the HCT 200 Metal Prep Set, select the 
Total Iron alternate form).

6. Wipe the outside of the zero vial. 

Place the vial into the adapter. 
Place the cover on the adapter.

7. Press: 

ZERO

. The display will show: 

Zeroing

… then: 

0.00 mg/L Free Fe 

or 

0.00 mg/L Total Fe 

if the Total Iron alternate form 

was selected.

8. Clean the outside of the sample vial.
9. Place the vial into the adapter. Place the cover on the 

adapter.

10. Press: 

READ

. The display will show: 

Reading

...

then the results are displayed in 

mg/L Free Fe 

or 

Total Fe

Note

If the test overranges, repeat the 

UniCell™

test with diluted

sample.

DR/2010

Note

If this is the first time you have run this test on your DR/890,
you must first perform the DR/890 User Calibration procedure
at the end of the next section (see Instrument Methods).

INSTRUMENT SET-UP

1. Press 

PRGM

. The display will show 

PRGM ?

Enter the user program number that the 

UniCell™

Iron HCT 159

was entered in and 

press 

ENTER

. The display will show 

USER

and 

mg/L

as 

the units. These will be in the Concentration Display 
Form unit chosen when the method was entered in the 
instrument.

2. Insert the COD/TNT Adapter into the cell holder by 

rotating the adapter until it drops into place.
Then, push down to make sure it is completely inserted.

3. Follow the procedure steps.

(

Note:

Ensure that the correct Iron unit Free or Total was 

used when the method was entered in the instrument).

4. Clean the outside of the Zero vial and place it in the 

adapter. Push down on the top of the vial until it seats 
completely into the adapter. Tightly cover the vial with 
the instrument cap and press 

ZERO

.

5. Clean the outside of the sample vial and place it in the 

vial adapter. Push down on the top of the vial until it 
seats completely into the adapter. Tightly cover the vial 
with the instrument cap and press 

READ

.

6. Results will be in 

mg/L

of the Display Form unit chosen 

when the method was entered.

DR/890

HCT 159

HCT 159

MEASURING RANGE: 

0.10 - 5.00 mg/L Free Fe / 0.12 - 6.00 mg/L Total Fe

MEASURING RANGE: 

0.10 - 5.00 mg/L Free-Fe / 0.12 - 6.00 mg/L Total-Fe

Note

If this is the first time you have run this revision of the test on
your DR/2500, you must contact your local 

Hach

distributor

for an upgrade of your software (Instrument Version Number
1.11 and Hach Program version number 27).

INSTRUMENT SET-UP

1. Touch: 

Hach Programs

.

Select program 

276

Iron HCT 159

.

Then touch: 

Start

.

2. Follow the procedure steps.

(

Note:

If measuring Total Iron and performed the sample 

digestion using the HCT 200 Metal Prep Set, select the 
Total Iron alternate form under Options).

3. Wipe the outside of the zero vial. 

Place it into the cell holder.

4. Touch: 

Zero

. The display will show: 

0.00 mg/L Free-Fe 

or 

Tot-Fe 

if the Total Iron 

alternate form was selected.

5. Clean the outside of the prepared sample vial and place 

it into the cell holder. Read the results.

DR/2500

HCT 159

MEASURING RANGE: 

0.10 - 5.00 mg/L Free-Fe / 0.12 - 6.00 mg/L Tot-Fe