background image

ZERO

5

1

3

4.0

ml

Sample

2

4

15

min

PROCEDURE 

for Free Iron

applies to DR/4000, DR/2500 DR/2010 and DR/800 

Note

If this is the first time you have run this revision of the test on
your DR/4000, you must contact your local 

Hach

distributor

for an upgrade of your software (Cat.No. 49544-00). Software
Version number 2.39 contains the factory 

UniCell™ 

programs.

INSTRUMENT SET-UP

1. Press the soft key under 

HACH PROGRAM 

(or 

USER PROGRAM

if the method was entered as a 

USER PROGRAM

). 

Enter program number 

2130 

(or 

159 

if the method was 

entered as a 

USER PROGRAM

).

Press 

ENTER

.

2. The display will show:

HACH PROGRAM: 2130
Iron, HCT 159

The wavelength (

λ

), 

485 nm

, is automatically selected.

3. Insert the 

Test N’Tube

adapter into the cell holder by

sliding it under the thumb screw and into the alignment
grooves. Fasten in place by tightening the thumb screw.

4. Follow the procedure steps.

(

Note:

If measuring Total Iron and performed the sample 

digestion using the HCT 200 Metal Prep Set, select the 
Total Iron alternate form under the Options soft key menu).

5. Wipe the outside of the zero vial.

Place the zero vial into the adapter. 
Close the light shield.

6. Press the soft key under 

ZERO

.

The display will show: 

0.00 mg/L Free-Fe 

or 

0.00 mg/L Tot-Fe 

if the Total Iron alternate form 

was selected.

7. Clean the outside of the sample vial.
8. Place the prepared sample vial into the adapter. Close 

the light shield. Results in 

mg/L Free-Fe

or 

Tot-Fe

are 

displayed. 

Note

If the test overranges, repeat the 

UniCell™

test with diluted

sample.

2

. Pipet 

4.0 ml

sample  

into the sample vial.

3

. Close sample vial and 

invert a few times, until 
the freeze-dried 
contents are 
completely dissolved.

4

. Start the instrument 

timer. 

15-minute

reaction 

period will beginn.

5

. Zero the instrument 

using the zero vial 
(white cap) provided 
and read the sample.

DR / 4000—Rev 2.39

1

. Enter program number

and set up the 
instrument as explained
for each instrument 
type.

HCT 159

HCT 159

LIST OF REQUIRED REAGENTS

Iron-Fe  

UniCell™

HCT 159

23/pkg

HCT 159

OPTIONAL REAGENTS

Metal Prep Set  

UniCell™

HCT 200

50 digestions HCT 200

Iron Standard 
1000 mg/L as Fe

100 ml

2271-42

Sodium Hydroxide
Standard Solution

1000 ml

2450-53

(5.0 N) 

Nitric Acid, concentrated

500 ml

152-49

OPTIONAL APPARATUS

Cylinder, mixing – 100 ml

each

20886-42

Flask, volumetric 100 ml

each

547-42

Pipettor, (Jencons) 1-5 ml

each

27951-00

Replacement tips for 27951-00

pk/100

27952-00

Pipettor, (Jencons) 100-1000 µl

each

27949-00

Replacement tips for 27949-00

pk/400

27950-00

pH Paper

pk/100

26013-00

MEASURING RANGE: 

0.10 - 5.00 mg/L Free-Fe / 0.12 - 6.00 mg/L Tot-Fe

PROCEDURE 

for Total Iron:

After digestion with the Metal Prep Set HCT 200,
follow the procedure steps 1 – 5.