Hach HQ11d Скачать руководство пользователя страница 25

4.

Insert the USB storage device that contains the software update files
into the USB/DC adapter.

5.

Turn on the meter.
The update process starts. The display will show “Updating meter to
<firmware version>”. After an interval, the display changes to
“Updating files, please wait...” In addition, the display will show a
rotating flask and emit a periodic audio signal during the update
process.

Note: A large capacity USB storage device increases the time required for
completion of the update process, even if most of the device memory is empty.

6.

Wait for the meter to finish the software update. When the update
process is complete, the message “Update complete. Remove USB
device” is shown. The meter will turn off after the USB device has
been removed.

7.

Repeat steps 1 through 6 to update the software in other HQd
meters as necessary.

Download software updates

To download the most current version of the software, refer to the
applicable product page on the manufacturer's website.

1.

Transfer the update files to a USB storage device.

2.

Follow the instructions in 

Update the meter software

 on page 22 to

update the software in the meter.

Transfer method settings

Probe settings that have been changed by the user for measurements,
calibrations or check standards (Meter Options > (Probe Model) Settings
> Modify Current Settings) can be copied to a USB storage device. The
USB device can then be used to transfer the method settings to other
HQd meters that accept the same probes.

N O T I C E

 

Make sure the USB storage device does not contain HQd meter software update
files to prevent unintentional updates.

1.

Turn off the meter.

2.

Connect the meter to AC power (

Figure 4

 on page 9).

3.

Plug the USB storage device into the USB/DC adapter before the
meter is powered on.

4.

Turn on the meter.

5.

Push the 

 key and select Transfer Methods. If the USB device

already contains a method settings file, an option to export or import
methods is shown. Select Export Methods.

6.

In the Select Methods to Export screen, select one or more methods
to copy to the USB device. A check mark is shown next to each
selected method.

7.

Select OK. The settings are copied to the USB storage device. When
complete, the Transfer Summary screen is shown.

8.

Connect the AC power and USB device to a meter that will receive
the method settings. Turn the meter on.

9.

Push the 

 key and select Transfer Methods. If the USB device

already contains a method settings file, an option to export or import
methods is shown. Select Import Methods.

10.

In the Select Methods to Import screen, select one or more methods
to transfer to the meter. A check mark is shown next to each selected
method.

11.

Select OK. The user method settings are transferred from the USB
storage device to the meter. When complete, the Transfer Summary
screen is shown. Select details to view additional information about
the transfer.

12.

Disconnect the USB storage device from the meter.

Bi-directional Communication between the meter and
a PC

For measurement automation the meter can be used to implement a
command set for meter remote control or automated data transfer. The
command set can be used to perform minimal configuration and to
control the meter. To set up the meter for communication and control,
refer to 

About meter configuration

 on page 24. For additional

information and the command set contact Technical Support.

About meter control

The virtual serial connection can be used to control meter functions from
a PC. For example, the functions include starting a measurement cycle,

English

   

23

Содержание HQ11d

Страница 1: ...DOC022 53 80017 HQd Portable Meter 10 2017 Edition 6 User Manual...

Страница 2: ......

Страница 3: ...er 8 User interface and navigation 9 User interface 9 Display description 9 Navigation 11 Startup 11 Turn the meter on and off 11 Change the language 11 Change the date and time 11 Connect a probe 12...

Страница 4: ...20 Set the sound options 20 Set the probe calibration reminder 21 Change the temperature units 21 Set the measurement mode 21 Set auto measurement intervals 21 Start interval measurements 21 Prevent...

Страница 5: ...Replacement parts and accessories 26 Examples of printed reports 28 Table of Contents 3...

Страница 6: ...Table of Contents 4...

Страница 7: ...ated USB type B for PC Temperature correction Off automatic and manual parameter dependent Measurement display lock Continuous measurement Interval or Press to Read mode Averaging function for LDO pro...

Страница 8: ...to the manufacturer for disposal at no charge to the user Product overview The HQd series portable meters are used with digital IntelliCAL probes to measure various parameters in water The meter autom...

Страница 9: ...the batteries and dry the interior of the compartment Check the battery contacts for corrosion and clean them if necessary N O T I C E When using nickel metal hydride NiMH batteries the battery icon...

Страница 10: ...ALWAYS be provided with a Ground Fault Circuit Interrupting GFCI GFI circuit breaker The AC DC power adapter for this product is not sealed and must not be used on wet benches or in wet locations with...

Страница 11: ...ange the reading screen view 2 OPERATOR ID associate data with an individual 7 GREEN RIGHT key reads selects confirms or stores data 3 BACKLIGHT illuminate the display screen 8 DATA LOG recall or tran...

Страница 12: ...can also be selected in the Display Options menu Refer to Set the display options on page 20 Figure 7 Single screen display big screen mode 1 Calibration status indicator 5 Main measurement unit 2 In...

Страница 13: ...r change a value 3 Push the GREEN RIGHT key to advance to the next space 4 Push the GREEN RIGHT key under OK to accept the value Startup Turn the meter on and off Push the key to turn on or turn off t...

Страница 14: ...ay shows the current time and date Note The time stamp for a probe is set when the probe is first connected to the meter This time stamp makes it possible to record the probe history and record the ti...

Страница 15: ...New Sample ID Enter a name for a new sample ID Delete Sample ID Delete an existing sample ID Use an operator ID The operator ID tag associates measurements with an individual operator All stored data...

Страница 16: ...nd select View Probe Data The current calibration and calibration history for the probe can be viewed Option Description View Current Calibration The current calibration information shows the calibrat...

Страница 17: ...s in the basic report Total report Three lines of data The first two lines contain the same information as in the advanced report Send data to a USB storage device N O T I C E The transfer of a large...

Страница 18: ...mat Time Time of reading stored in user defined format Operator ID Operator ID that was used when the data was recorded Shows if no operator ID is used Table 1 Spreadsheet column descriptions continue...

Страница 19: ...fficient without a unit may be blank Table 1 Spreadsheet column descriptions continued Column header name Data description and example values Cal Number of Std s Number of standards used during calibr...

Страница 20: ...e display to show Transfer Complete The data is sent as a comma separated values csv file The data is shown in the HQ40d PC Application window Advanced operation Security options The Security Options...

Страница 21: ...ettings are used Table 2 Full Access Options Option Description Probe model settings Settings such as measurement options calibration options check standard options units and resolution Refer to the p...

Страница 22: ...ot active when the meter is connected to AC power or in the Interval Reading Mode Backlight The display backlight is turned off when the key is pushed Is it possible to set a time period after which t...

Страница 23: ...matically when the stability criteria are met Interval The sample is measured at regular intervals for a specified duration refer to Set auto measurement intervals on page 21 Data is stored in the dat...

Страница 24: ...nt information menu shows specific information such as the serial number for the meter or IntelliCAL R probe s 1 Push the key and select Instrument Information 2 Select Probe model Information or Mete...

Страница 25: ...n page 9 3 Plug the USB storage device into the USB DC adapter before the meter is powered on 4 Turn on the meter 5 Push the key and select Transfer Methods If the USB device already contains a method...

Страница 26: ...ia floppy CD ROM option and select the Include this location in the search and click the Browse button 15 Select the INF folder or location and click OK 16 Click Next The new software will be installe...

Страница 27: ...nt version of the software refer to the applicable product page on the manufacturer s website Problem with probe Connect a different IntelliCAL probe to verify if problem is with probe or meter Probe...

Страница 28: ...ebsite Error Warning Description Solution Meter will not power on or powers on intermittently Batteries are not installed correctly Examine battery orientation to make sure the batteries follow the po...

Страница 29: ...nsert for meter and probe storage 4 containers for sample collection 8505501 Field Case for 2 probes with greater than 5 m cables 30 m total Includes empty case insert for meter with protective glove...

Страница 30: ...e 12 Figure 12 Report header 1 Meter serial number1 3 Date and time 24 h YYMMDDhhmm 2 Report label 4 File type extension Data reports Sample data can be printed in a basic advanced or total report for...

Страница 31: ...DD MM YY or user defined format 9 Operator ID user defined shows if undefined 5 Measured value Figure 14 Advanced report for sample data 2 lines 1 Probe model 4 Additional units shows all additional u...

Страница 32: ...d reports Check standard data is printed with 1 line of information Figure 16 Figure 16 Check standard report 1 Report type ST check standard 5 Measured value 2 Report type ST check standard 6 Units 3...

Страница 33: ...parameter and user settings May be blank 4 Time of calibration prefaced by CAL and displayed as hh mm in 24 h or user defined format 10 r contents vary depending on type of parameter being read confi...

Страница 34: ...arameter pH LDO etc 6 Probe serial number a or on the HQ40d meter indicates the probe position 3 Time of calibration prefaced by CAL and displayed as hh mm in 24 h or user defined format 7 Calibration...

Страница 35: ......

Страница 36: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1...

Отзывы: