Table 3  Operator access options

Option

Description

(Probe model) settings Only methods (if methods exist) can be selected. Refer

to the probe documentation.

Note: A probe must be connected to the meter.

Run check standard

Measure standard solution (available for pH,
conductivity, ORP and ISE probes)

Note: A probe must be connected to use this option.

Instrument information Probe information

Meter information

Access password

Enter password

Display options

Contrast

Auto shutoff

Backlight

Mode

Sounds

Key press

Stability alert

Calibration reminder

Set the display options

Use Display Options to change the display contrast, battery saving auto-
shutoff options, the backlight option or the detailed or big reading screen
mode.

1.

Push the   key and select Display Options.

2.

Select which display option to change.

Option

Description

Contrast

Adjust the contrast of the display. The lightest setting is 0 and
the darkest setting is 9.

Option

Description

Auto-
shutoff

To maximize battery life, set a time period after which the
meter will automatically power off if no key is pushed (1, 2, 5,
10, 30 min, 1 h, 2 h or never). Auto-shutoff is not active when
the meter is connected to AC power or in the Interval Reading
Mode.

Backlight

The display backlight is turned off when the 

 key is pushed.

Is it possible to set a time period after which the backlight will
automatically power off if no key is pushed.

Mode

Select Detailed or Big screen size. Detailed will show more
information with smaller numbers and text. Big will show less
information with larger numbers and text.

Note: The screen size can also be selected from the measure
mode (refer to 

Display description

 on page 9).

Set the sound options

The meter can make an audible sound when a key is pushed, when
stability is reached or when the calibration reminder is due. The meter
also makes an audible sound when it begins transferring data to a USB
storage device and again when the data transfer is complete.

1.

Push the   key and select Sound.

2.

Choose which events will produce an audible sound. Multiple items
can be selected.

Option

Description

Key Press

The meter will make an audible sound whenever a key is
pushed.

Stability Alert

The meter will make an audible sound whenever
measurement stability is reached.

Cal reminder

The meter will make an audible sound when calibration is
due.

Note: Refer to 

Set the probe calibration reminder

on page 21 to set the calibration reminder to on or off.

20

   

English

Содержание HQ11d

Страница 1: ...DOC022 53 80017 HQd Portable Meter 10 2017 Edition 6 User Manual...

Страница 2: ......

Страница 3: ...er 8 User interface and navigation 9 User interface 9 Display description 9 Navigation 11 Startup 11 Turn the meter on and off 11 Change the language 11 Change the date and time 11 Connect a probe 12...

Страница 4: ...20 Set the sound options 20 Set the probe calibration reminder 21 Change the temperature units 21 Set the measurement mode 21 Set auto measurement intervals 21 Start interval measurements 21 Prevent...

Страница 5: ...Replacement parts and accessories 26 Examples of printed reports 28 Table of Contents 3...

Страница 6: ...Table of Contents 4...

Страница 7: ...ated USB type B for PC Temperature correction Off automatic and manual parameter dependent Measurement display lock Continuous measurement Interval or Press to Read mode Averaging function for LDO pro...

Страница 8: ...to the manufacturer for disposal at no charge to the user Product overview The HQd series portable meters are used with digital IntelliCAL probes to measure various parameters in water The meter autom...

Страница 9: ...the batteries and dry the interior of the compartment Check the battery contacts for corrosion and clean them if necessary N O T I C E When using nickel metal hydride NiMH batteries the battery icon...

Страница 10: ...ALWAYS be provided with a Ground Fault Circuit Interrupting GFCI GFI circuit breaker The AC DC power adapter for this product is not sealed and must not be used on wet benches or in wet locations with...

Страница 11: ...ange the reading screen view 2 OPERATOR ID associate data with an individual 7 GREEN RIGHT key reads selects confirms or stores data 3 BACKLIGHT illuminate the display screen 8 DATA LOG recall or tran...

Страница 12: ...can also be selected in the Display Options menu Refer to Set the display options on page 20 Figure 7 Single screen display big screen mode 1 Calibration status indicator 5 Main measurement unit 2 In...

Страница 13: ...r change a value 3 Push the GREEN RIGHT key to advance to the next space 4 Push the GREEN RIGHT key under OK to accept the value Startup Turn the meter on and off Push the key to turn on or turn off t...

Страница 14: ...ay shows the current time and date Note The time stamp for a probe is set when the probe is first connected to the meter This time stamp makes it possible to record the probe history and record the ti...

Страница 15: ...New Sample ID Enter a name for a new sample ID Delete Sample ID Delete an existing sample ID Use an operator ID The operator ID tag associates measurements with an individual operator All stored data...

Страница 16: ...nd select View Probe Data The current calibration and calibration history for the probe can be viewed Option Description View Current Calibration The current calibration information shows the calibrat...

Страница 17: ...s in the basic report Total report Three lines of data The first two lines contain the same information as in the advanced report Send data to a USB storage device N O T I C E The transfer of a large...

Страница 18: ...mat Time Time of reading stored in user defined format Operator ID Operator ID that was used when the data was recorded Shows if no operator ID is used Table 1 Spreadsheet column descriptions continue...

Страница 19: ...fficient without a unit may be blank Table 1 Spreadsheet column descriptions continued Column header name Data description and example values Cal Number of Std s Number of standards used during calibr...

Страница 20: ...e display to show Transfer Complete The data is sent as a comma separated values csv file The data is shown in the HQ40d PC Application window Advanced operation Security options The Security Options...

Страница 21: ...ettings are used Table 2 Full Access Options Option Description Probe model settings Settings such as measurement options calibration options check standard options units and resolution Refer to the p...

Страница 22: ...ot active when the meter is connected to AC power or in the Interval Reading Mode Backlight The display backlight is turned off when the key is pushed Is it possible to set a time period after which t...

Страница 23: ...matically when the stability criteria are met Interval The sample is measured at regular intervals for a specified duration refer to Set auto measurement intervals on page 21 Data is stored in the dat...

Страница 24: ...nt information menu shows specific information such as the serial number for the meter or IntelliCAL R probe s 1 Push the key and select Instrument Information 2 Select Probe model Information or Mete...

Страница 25: ...n page 9 3 Plug the USB storage device into the USB DC adapter before the meter is powered on 4 Turn on the meter 5 Push the key and select Transfer Methods If the USB device already contains a method...

Страница 26: ...ia floppy CD ROM option and select the Include this location in the search and click the Browse button 15 Select the INF folder or location and click OK 16 Click Next The new software will be installe...

Страница 27: ...nt version of the software refer to the applicable product page on the manufacturer s website Problem with probe Connect a different IntelliCAL probe to verify if problem is with probe or meter Probe...

Страница 28: ...ebsite Error Warning Description Solution Meter will not power on or powers on intermittently Batteries are not installed correctly Examine battery orientation to make sure the batteries follow the po...

Страница 29: ...nsert for meter and probe storage 4 containers for sample collection 8505501 Field Case for 2 probes with greater than 5 m cables 30 m total Includes empty case insert for meter with protective glove...

Страница 30: ...e 12 Figure 12 Report header 1 Meter serial number1 3 Date and time 24 h YYMMDDhhmm 2 Report label 4 File type extension Data reports Sample data can be printed in a basic advanced or total report for...

Страница 31: ...DD MM YY or user defined format 9 Operator ID user defined shows if undefined 5 Measured value Figure 14 Advanced report for sample data 2 lines 1 Probe model 4 Additional units shows all additional u...

Страница 32: ...d reports Check standard data is printed with 1 line of information Figure 16 Figure 16 Check standard report 1 Report type ST check standard 5 Measured value 2 Report type ST check standard 6 Units 3...

Страница 33: ...parameter and user settings May be blank 4 Time of calibration prefaced by CAL and displayed as hh mm in 24 h or user defined format 10 r contents vary depending on type of parameter being read confi...

Страница 34: ...arameter pH LDO etc 6 Probe serial number a or on the HQ40d meter indicates the probe position 3 Time of calibration prefaced by CAL and displayed as hh mm in 24 h or user defined format 7 Calibration...

Страница 35: ......

Страница 36: ...h com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1...

Отзывы: