Operator Manual
4.4
Loading Samples with the Large Volume Adapter
Attach the sample vessel to the 9705 as described in
“Connecting the Large Volume Adapter”
. Make sure the stir speed is set to 0 (off).
4.5
Running a Sample
1) Once the bottle is loaded or the large volume adapter is attached, double-check that
the settings in the upper right hand corner of the display are correct. If necessary,
change settings as described in
“Changing Sample Parameters” on page 49
.
2) From the
Main Screen
,
press the
Swirl
icon to begin particle dispersion.
3) Press the
Run
icon to start the sample run. A red light appears to the right of the
Lock
icon in the icon bar, as shown in
. This light appears when a sample is being
drawn, during both tare volume and sample volume.
4) While the sample run is in progress, parameters cannot be changed. Click on the lower
portion of the main screen to toggle between viewing the data in a chart form and in a
graph form. In the chart form, use the blue arrows at the right of the screen to scroll
through the data. Runs appear on Y-axis; particle sizes appear on the X-axis.
5) If laser current or calibration is detected to be out of range, a yellow caution triangle
appears on the screen. Click on the triangle to invoke the
Diagnostic
screen. Refer to
“Diagnostics and Troubleshooting” on page 65
for more information.
6) When the run is finished and all samples have been taken from the bottle, the red light
will disappear. Tilt the bottle dock away from the unit to begin unloading the sample.
7) Place fingers through the holes on the side of the bottle dock. Gently but firmly pull up
on the bottle, being careful not to come in contact with the needles.
8) Dispose of any sample bottles according to standard operating procedures (SOPs). Do
not re-use samples since they are no longer sterile.
Fig 4-2 :
Main Screen with Red Status Light
Test Running Indicator
Содержание HIAC 9705
Страница 1: ...HIAC 9705 Liquid Particle Counting System Operator Manual...
Страница 2: ...9705 KSS March 2006 Edition 1 Operator Manual...
Страница 6: ...6 of 82 Table of Contents 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 10: ...10 of 82 Manual Overview 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 22: ...22 of 82 Initialization 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 60: ...60 of 82 Operating the 9705 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 64: ...64 of 82 Maintenance Procedures 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 72: ...72 of 82 Service Procedures 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Страница 78: ......
Страница 81: ...9705 User Notes 81 of 82 KSS March 2006 Edition 1 HIAC Operator Manual User Notes...