FlowCell FC 48/10 USC
10 English
To trigger a cleaning process there are two options: You can use the 5 V signal that
is led out of the FlowCell to realize a triggering over an external switch contact.
Alternatively an external signal voltage of 5 to 24 V DC (in reference to GND-line)
can be put on the “Trigger input” line.
1.4.5 Connection of power supply
The power supply is connected through the power socket of the FlowCell like it is
shown in the following polarity illustration.
WA R N I N G
Electrical shock and fire hazards. Make sure that the supplied cord and non-locking plug
meet the applicable country code requirements.
N O T I C E
The manufacturer recommends to use only the supplied power supply.
N O T I C E
For the power supply of the FlowCell electronics, a power supply unit of 12 to 24 V DC with
an output cable with at least 15 W is needed.
FlowCell
External measurement
controller
Cleaning cycle
Trigger
μC
DC
DC
blue
white
brown
FlowCell
External measurement
controller
Cleaning cycle
Trigger
+5..24 VDC
GND
μC
DC
DC
blue
white
brown
2,1 x 5,5
–
+
Содержание Flow Cell FC 48/10 USC
Страница 3: ...Table of Contents 2...
Страница 13: ...FlowCell FC 48 10 USC 12 English...
Страница 23: ...FlowCell FC 48 10 USC 22...
Страница 33: ...FlowCell FC 48 10 USC 32...
Страница 43: ...FlowCell FC 48 10 USC 42...
Страница 53: ...FlowCell FC 48 10 USC 52...
Страница 63: ...FlowCell FC 48 10 USC 62...
Страница 73: ...FlowCell FC 48 10 USC 72...
Страница 83: ...FlowCell FC 48 10 USC 82...
Страница 93: ...FlowCell FC 48 10 USC 92...
Страница 103: ...FlowCell FC 48 10 USC 102...