4.
Select Setup Port [1] (sensor name).
5.
Complete the options in the Basic Settings menu.
Option
Description
AC Frequency
Selects the power line frequency to get the best noise rejection. Options: 50 or 60 Hz
(default).
Calibrate the sensors
Calibrate the sensors during sensor setup and at regular intervals. Make sure that the sensor is
installed in the process before this task is started.
1.
Push
MENU
and select Calibration.
2.
Select the sensor to calibrate.
3.
Obey the screen prompts to select the necessary values.
4.
Wait for the screen to show Calibration Complete with a summary of the calibration data.
Note: If the calibration fails, make sure that the parameters in the Sensor Setup menu are correct.
5.
Select Finish. The Verify screen shows.
6.
Select Yes to take a measurement to verify the calibration.
7.
Examine the measurement data to determine if the measurement is correct.
Configure data logging
N O T I C E
All data and alarm logs for all channels are erased in the logger when channels are added to or removed from a
program. Make sure to download the data from the logger to a safe location first, then change the program.
Use the Programming menu to set the channels to record in the data log. A channel can be a reading
from an attached sensor (e.g., level, flow, temperature), the battery voltage from a backup battery or
a statistical value based on a sensor reading. Data logging starts only when a channel is selected.
1.
Push
MENU
.
2.
Select Programming>Datalog Programming.
3.
Select Channels Logging.
4.
Select the sensor or logger.
5.
Select a maximum of 16 channels.
Note: The port number where the sensor connects to the instrument shows in the sensor channel name. For
example, Velocity 2 is the velocity channel name for the sensor that connects to sensor port 2. An IO number
that follows a channel name identifies the analog inputs of the optional I/O module.
6.
Select Save.
7.
Select Back>Logging Intervals>sensor or logger to set the logging interval.
8.
Select the primary and secondary logging interval. The primary logging interval operates during
normal operation. The secondary logging interval operates during alarm conditions.
Note: The primary and secondary logging intervals operate for all channels of a sensor or the logger. A logging
interval for an individual channel is not possible.
Configure the alarms
N O T I C E
All alarm logs are erased when programmed alarms are deleted. Make sure to save the settings and data first,
then change the program.