c.
Put the vial cover on the sample vial.
d.
Push
Enter
to select
READ
.
10.
Measure the Standard 2 vial as follows:
a.
Clean the Standard 2 vial with a no-lint cloth.
b.
Install the Standard 2 vial in the vial adapter.
c.
Put the vial cover on the sample vial.
d.
Push
Enter
to select
READ
.
11.
Identify if a slope calibration is necessary as follows:
•
Total chlorine
—If the Standard 1 reading is not within ±2 ppb of
the total chlorine value from the Certificate of Analysis and the
Standard 2 reading is not within ±5 ppb of the total chlorine value
from the Certificate of Analysis, do a slope calibration for total
chlorine.
•
Free chlorine
—If the Standard 1 reading is not within ±2 ppb of
the free chlorine value from the Certificate of Analysis and the
Standard 2 reading is not within ±5 ppb of the free chlorine value
from the Certificate of Analysis, do a slope calibration for free
chlorine.
•
Sulfite
—If the Standard 1 reading is not within ±5 ppb of the
sulfite value from the Certificate of Analysis and the Standard
2 reading is not within ±20 ppb of the sulfite value from the
Certificate of Analysis, do a slope calibration for sulfite.
8.3 Do a slope calibration
Do a 2-point slope calibration only if the instrument does not pass a
calibration check.
1.
Push
Enter
to set the instrument to on. Keep the sample port cover
closed during startup.
2.
After startup, make sure that the sample port is clean. Use a soft
cloth or no-lint paper tissue (or air duster) to clean the sample port
as necessary.
3.
Install the vial adapter in the instrument. Push in the tabs on the
sides of the vial adapter.
4.
Select
CALIBRATION
, then select the parameter for the
calibration.
18
English
Содержание DR1300 FL
Страница 2: ...Table of Contents English 3 Espa ol 27 Fran ais 54 Portugu s 81 108 129 153 2...
Страница 110: ...3 1 1 3 1 2 3 1 3 IC Canadian Radio Interference Causing Equipment Regulation ICES 003 B B FCC 15 B 110...
Страница 111: ...FCC 15 1 2 FCC 15 B 1 2 3 3 2 Hach DR1300 FL DR1300 FL DR1300 FL DR1300 FL 3 3 2 DR1300 FL 1 DR1300 FL 111...
Страница 112: ...1 DR1300 FL 1 3 3 2 4 LCD 3 112...
Страница 113: ...3 4 2 2 1 DR1300 FL 4 2 5 USB 4 3 2 6 AA 4 113...
Страница 114: ...4 AA 3 3 114...
Страница 115: ...5 5 1 4 4 1 Enter 5 4 Right 2 Left 5 Down 3 Up 115...
Страница 116: ...5 2 5 5 1 5 2 6 3 7 4 116...
Страница 118: ...7 MSDS SDS 119 1 Enter 2 3 a b c d 119 e f g 4 118...
Страница 119: ...7 1 7 2 1 http www hach com 2 DR1300 FL 3 Search Type Downloads 4 Methods Procedures 5 8 MSDS SDS 119...
Страница 124: ...10 10 1 AA 114 10 2 10 3 124...
Страница 125: ...11 120 120 120 120 125...
Страница 126: ...124 120 120 120 DR1300 FL DPD ppb 12 126...
Страница 128: ...1 3563500 16 mm 6 pkg 100866 128...
Страница 131: ...3 1 3 1 1 131...
Страница 133: ...1 2 3 3 2 Hach DR1300 FL DR1300 FL DR1300 FL DR1300 FL 3 3 2 DR1300 FL 1 DR1300 FL 133...
Страница 134: ...1 DR1300 FL 1 3 3 2 4 LCD 3 134...
Страница 135: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB 4 3 2 6 AA 4 4 Bluetooth 135...
Страница 136: ...4 AA 4 3 3 136...
Страница 137: ...5 5 1 4 4 1 Enter 5 4 2 5 3 137...
Страница 138: ...5 2 5 5 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 138...
Страница 140: ...Update Program Factory Default Factory Exit 7 MSDS SDS 141 1 Enter 2 TESTS 140...
Страница 141: ...3 a b TIMER START c d 141 e f g BACK READ 4 7 1 7 2 1 http www hach com 2 DR1300 FL 3 141...
Страница 142: ...4 Methods Procedures 5 8 MSDS SDS 8 1 1 Enter 2 3 141 4 CALIBRATION 5 CAL RB 6 a b c d e Enter 142...
Страница 147: ...LOW BATTERY 4 AA 136 10 2 10 3 147...
Страница 148: ...11 SYSTEM Diagnosis Sample Collection 142 143 148...
Страница 149: ...142 142 147 142 142 143 149...
Страница 150: ...DR1300 FL Sample Collection DPD ppb 12 150...
Страница 152: ...1 3563500 16mm 6 100866 152...
Страница 155: ...3 1 1 3 1 2 3 1 3 ICES 003 155...
Страница 157: ...DR1300 FL 1 DR1300 FL 1 DR1300 FL 1 3 3 Keypad 2 4 LCD 3 157...
Страница 158: ...3 4 2 2 1 DR1300 FL 4 2 5 Bluetooth USB4 3 2x 6 AA 4 Bluetooth 158...
Страница 159: ...4 AA 3 3 159...
Страница 160: ...5 5 1 Keypad 4 4 1 Enter 5 4 2 5 3 160...
Страница 161: ...5 2 Display 5 5 Main menu 1 SYSTEM 5 2 TESTS 6 CALIBRATION 3 7 POWER 4 161...
Страница 163: ...Factory Default Factory Exit 7 MSDS SDS 164 1 Enter 2 TESTS 3 a 163...
Страница 164: ...b TIMER START c d 164 e f g BACK READ 4 7 1 7 2 1 http www hach com 2 DR1300 FL 3 4 Methods Procedures 5 164...
Страница 165: ...8 MSDS SDS 8 1 1 Enter 2 3 164 4 CALIBRATION 5 CAL RB 6 a b c d e Enter 7 Enter 8 165...
Страница 170: ...11 SYSTEM Diagnosis 165 166 165 170...
Страница 171: ...165 169 165 165 166 DR1300 FL DPD 171...
Страница 172: ...12 DR1300 FL 34630000 16 34252000 16 34251000 16 34250000 100 34252001 100 34251001 100 34250001 172...
Страница 174: ......
Страница 175: ......