Table of contents
on page 3
on page 18
on page 20
on page 7
on page 16
on page 21
on page 17
Legal information
Manufacturer: TriOS Mess- und Datentechnik GmbH
Distributor: Hach Lange GmbH
The translation of the manual is approved by the manufacturer.
Specifications
Specifications are subject to change without notice.
Specification
Details
Dimensions (W x D x H)
280 × 94 × 170 mm (11 × 3.7 × 6.7 in.)
Weight
3.7 kg (8.16 lb)
Enclosure
IP66 die-cast aluminium alloy housing; Front panel: acrylic glass (PMMA)
Operating temperature
0 to 40 °C (–4 to 140 °F)
Storage temperature
–20 to 70 °C (–4 to 158 °F)
Relative humidity
0 to 95%, non-condensing
Power requirements
85–265 VAC, 50/60 Hz; 12–24 VDC ± 5%
Power consumption
6 W (50 W maximum)
Measurement
Four M12 connectors for optical sensors
Memory
Internal 2 GB Micro SD card
Communications
RS232, RS485
Communications protocol Modbus RTU: RTU server, RTU client, configurable parameters (default: 9600-8-N-1)
Modbus TCP: TCP server, configurable TCP port (default: 502)
Network connection
Ethernet, RJ-45 connector, protocol: TCP/IP, Modbus TCP, VNC
USB port
USB 2.0 (Host), USB-A socket for data download
Relays
One relay (SPDT), 250 VAC, 2 A; Wire gauge: 2.5 mm
2
(14 AWG) maximum
Outputs
Six 4–20 mA configurable analog outputs; Load: 500
Ω
maximum; Error indicator:
0 mA; Wire gauge: 1.5 mm
2
(16 AWG) maximum
Input/output switch
Measurement trigger: Trigger for global measurement (electrically isolated); Control
voltage: 12–24 VDC (± 5%); Wire gauge: 1.5 mm
2
(16 AWG) maximum
Control voltage: no
Cleaning
Integrated compressed air valve for cleaning
Compliance information
CE mark
Warranty
1 year (EU: 2 years)
English
3
Содержание CD500
Страница 2: ...English 3 Deutsch 22 Italiano 43 Fran ais 64 Espa ol 85 Portugu s 106 e tina 127 Polski 146 2...
Страница 167: ......