2.2.5  Reaction Archive Screen 

The Reaction Archive holds information on TIC, TOC, TC, VOC (in mgC/l), COD, BOD (in mgO/l), Drinking 

Water parameters (REMOVAL in %, and RESULT in FAIL/PASS condition), stream valve, reaction range, 

start time and related analysis information for the last 9999 reactions. If the archive is full, then every new 

reaction overwrites the oldest one in the archive. As the Reaction Archive contains 9999  events, the user 

must first enter the date at which the viewing of the archive starts. The Enter Date menu allows the user to 

specify the date of the first displayed reaction from the archive. 

 

Each reaction record in the reaction archive contains:  

-  Start Time  

- reaction start time, which is displayed without seconds in this menu 

-  Date    

- reaction date 

-  Reaction Type  - with the prefixes below: 

 

S1 to S3: 

Reactions from stream 1 to stream 3. 

M1 to M3: 

Reactions from manual sample stream 1 to manual stream 3. 

 

 

√ 

 

Sample sensor detected the sample or there is no significant quantity of air 

 

 

 

 

bubbles in the stream/manual grab sample lines. 

Sample sensor detected  no  sample  or there is significant quantity of air 

bubbles in the stream/manual grab sample lines. See Sample Status in 

section 

8.3.8  Fault Setup

 for details. 

RS: 

 

Remote standby reaction. 

ZC: 

 

Zero calibration reaction. 

ZK: 

 

Zero check reaction. 

ZM: 

 

Manually input zero adjust. 

SC: 

 

Span calibration reaction. 

SK: 

 

Span check reaction. 

SM: 

 

Manually input span adjust. 

A1 to A3: 

24 hours average result from stream 1 to stream 3. 

 

The user can navigate through the displayed reactions individually by pressing the UP and DOWN keys each 

time, or can navigate in steps of 10 reactions using the LEFT and RIGHT keys. Depending on system 

analysis type (e.g. VOC, TC –TIC etc.) and system display options (e.g. COD and/or BOD) settings, BioTector 

displays additional reaction data held on additional Reaction Archive screens. To access the screens, press 

the ENTER key, and to return to the previous screen, press the ESCAPE key. 

2.2.6  Fault Archive Menu 

In the Fault Archive menu, the user can view the last 99 faults/warning/notification events logged in the 

system, confirm if these events are current or not, and acknowledge the current events. If the archive is full, 

then every new event overwrites the oldest one in the archive. The user can navigate through the displayed 

reactions individually by pressing the UP and DOWN keys each time, or can navigate in steps of 10 reactions 

using the LEFT and RIGHT keys. See 

Section 9  

Troubleshooting  of System Fault, Warning and 

Notification Events

 for a list of all systems fault, warning and notification events.  

 

The faults archive events are divided into three categories: 

-  Fault:

 Faults are categorized as events, which stop BioTector operation. The 4-20mA signals are set 

to the fault level, and the fault relay is activated. The BioTector cannot be started unless the fault in 

the archive has been acknowledged. 

-  Warning: 

Warning is a minor event, which does not require the BioTector to stop. The 4-20mA 

signals are not changed, only the fault relay is activated. 

-

 

Notification:

  A notification is an information (e.g. “86_Power Up”, “87_Service Time Reset” etc.) 

displayed on the screen. 

 

 

To acknowledge any current events  marked with an “*” in the archive, first identify and locate the 

faults/warnings/notification. Follow the necessary troubleshooting procedures to solve the problem. See 

Section 9 

 

Troubleshooting of System Fault, Warning and Notification Events 

for details. 

Acknowledge the fault by pressing the ENTER key in the Fault Archive menu. Please note that there are 

system faults (e.g. 05_Pressure Test Fail), which cannot be acknowledged by the user. Such faults are reset 

and acknowledged automatically by the system when system is started, when system is rebooted or when the 

fault condition is solved.  If an event cannot be acknowledged when the system is running, a “SYSTEM 

RUNNING” message is displayed on screen.  

Содержание BioTector B3500

Страница 1: ...DOC023 53 90740 BioTector B3500 Online TOC Analyzer BASIC USER MANUAL 6 2018 Edition 1...

Страница 2: ......

Страница 3: ...cautions 12 1 5 3 Pr cautions chimiques 12 1 5 4 Sample Stream Precautions 13 1 5 4 Pr cautions relatives aux chantillons 14 SECTION 2 OPERATOR S MANUAL 15 2 1 SOFTWARE SCREENS AND SOFTWARE MENU DIAGR...

Страница 4: ...be impaired 1 1 Information and Safety Signs used in the Manual When any supplementary information is required and if any hazards exist the necessary information and safety signs Information Caution...

Страница 5: ...sary to work with this component it should be handled with care This symbol when noted on a product illustrates the risk of chemical harm due to its corrosive acidic caustic or solvent nature Only qua...

Страница 6: ...rop ene indicates that The instrument complies with the European product directives health safety and environmental protection legislations Conforms to ANSI UL Std 61010 1 Certified to CAN CSA Std 610...

Страница 7: ...ned herein or for incidental or consequential damages in connection with the furnishing performance or use of this material The information contained in this manual is subject to change without notice...

Страница 8: ...ary oedema may develop with delayed onset usually one or more hours after exposure Following severe acute ozone toxification recovery is slow In the few severe human cases reported 10 14 days of hospi...

Страница 9: ...s g n rales de s curit Pri re d tre toujours attentif toutes les notices de prudence d avertissement ou de danger Le non respect des instructions de s curit peut engendrer la blessure grave d individu...

Страница 10: ...en Isoler les fils lectriques de l appareil avant de commencer tout travail dans le bo tier lectronique Seul le personnel lectricien qualifi est habilit effectuer tous travaux d lectricit Se conformer...

Страница 11: ...t Pour son fonctionnement BioTector emploie de l oxyg ne O2 comme gaz porteur L oxyg ne ne doit comporter aucun gaz carbonique CO2 ni d azote N2 Le taux moyen de consommation d oxyg ne du BioTector es...

Страница 12: ...sing appropriate safety equipment 1 5 3 Pr cautions chimiques La liste de la Section 6 R actifs et Standards de Calibration Section 6 Reagents and Calibration Standards num re un certain nombre de pro...

Страница 13: ...FA Per fluoro alkoxy Vinyl Reinforced Silicon Reinforced PVC Poly vinyl chloride Fittings PFA Per fluoro alkoxy Stainless Steel SS 316 PVDF Poly vinylidene flouride Brass Pump Tubing EMPP Elastomer mo...

Страница 14: ...xy Vinyle Silicone renforc PVC renforc poly chlorure de vinyle Raccords PFA perfluoroalkoxy Acier inoxydable SS 316 PVDF polyfluorure de polyvinylid ne Laiton Tuyauterie de la pompe PPMOD polypropyl n...

Страница 15: ...ramming the BioTector The ENTER key which advances user to the next screen is also used to enter programmed settings in the BioTector The symbols used on BioTector LCD screen and their meanings are as...

Страница 16: ...Page 16 Software Menu Diagram...

Страница 17: ...yed on the top left hand side of the Analysis Data and Reagent Status screens On most other screens only the screen name is displayed in this location System status messages are displayed in the follo...

Страница 18: ...mple if the reaction is currently in the TIC Base Oxidation TOC phase The operation Range e g Range 1 2 or 3 the BioTector is using to carry out its analysis The Reaction Time which is the elapsed tim...

Страница 19: ...he analyzer in C e g 26 C The CO2 instantaneous value e g 56ppm CO2i and the CO2 zero value e g 12ppm CO2z of the reaction The elapsed time e g 265s since the start of the analysis 2 1 5 Reagent Statu...

Страница 20: ...r diagnostics to download data to program the software functions and to program the system specific settings in the BioTector The sub menus in this level can be password protected using the Password m...

Страница 21: ...Standby in the system reaction archive The 4 20mA signal or other output devices are not updated When remote standby signal is deactivated the BioTector starts analyzing When remote standby signal is...

Страница 22: ...eyboard the BioTector stops as soon as its present reaction is completed An is displayed to let the operator know the function has been activated 4 Emergency Stop When this function is activated the B...

Страница 23: ...ils The comprehensive Zero Check Calibration and Span Check Calibration parameters are programmed in Zero Program and Span Program menus respectively see Section 8 3 4 2 Zero Program and 8 3 4 3 Span...

Страница 24: ...n Check menu 3 If Span Calibration or Span Check is activated in New Reagents Program menu a Span Calibration or Span Check is carried out Once the procedure is completed the BioTector either stops or...

Страница 25: ...cate that this function has been selected If the BioTector is stopped then the Manual mode will start immediately To deactivate this function before the manual operation mode has started press the ENT...

Страница 26: ...on Archive screens To access the screens press the ENTER key and to return to the previous screen press the ESCAPE key 2 2 6 Fault Archive Menu In the Fault Archive menu the user can view the last 99...

Страница 27: ...startup time for a scheduled task for example the startup time for a manual sample sequence in Manual Program menu 2 2 8 Contact Information Contact Information menu displays the manufacturer distribu...

Страница 28: ...em only checks the zero response at each range and displays the suggested Zero Adjust values in brackets for all ranges next to the current Zero Adjust settings When a Zero Check cycle is completed if...

Страница 29: ...8 Zero Program Zero Program function allows the user to program the number of zero reactions run at one or more ranges R1 R2 and or R3 When the number of zero calibration reactions for one or two of t...

Страница 30: ...below are displayed in this menu 1 3 TOC Span Adjust This menu item allows the user to set the TOC span adjust factors manually by entering the STANDARD solution used and the calibrated average react...

Страница 31: ...alibration and Span Check reactions are carried out If the selected range is in conflict with the programmed CALIBRATION STANDARD concentration in this menu system automatically displays a Caution Rea...

Страница 32: ...for TOC Calibration Standard above Section 8 3 4 3 Span Program and Section 6 2 Calibration Standards for further details 21 TC Check Std In VOC systems the sum of the TIC and TOC Check Standard solut...

Страница 33: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...

Страница 34: ...years The product manufacturing date can be found on the product label on the main board enclosure behind the display BioTector 25 EFUP CHINA ROHS DECLARATION TABLE ROHS Part name Toxic or Hazardous...

Страница 35: ......

Страница 36: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...

Отзывы: