For sensors connected to an FL1500 flow logger or AS950 sampler, refer to the FL1500 flow logger
or sampler documentation to configure the sensors. As an alternative, configure the sensors with the
FS
DATA Desktop Setup Wizard when the sensors are connected to an FL1500 flow logger.
For sensors connected to a Sigma flow meter, do the steps in
Level calibration for Sigma flow meters
on page 15.
Note: If a sensor is replaced, removed for maintenance or moved to another instrument, do a level calibration.
Level calibration for Sigma flow meters
1.
With the sensor installed in the flow, monitor the Current Status with a PC using Insight software
or a flow meter display.
2.
Physically measure the distance from the top of the pipe to the surface of the water. Refer to
3.
Subtract the number from step 2 from the pipe diameter. Refer to
.
The result is the water depth. Refer to
.
4.
Use the Adjust Level function of the software to enter the physically-measured water depth.
Figure 9 Measure the water level
1
Water level
Maintenance
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Clean the sensor
Clean the transducer port when:
• Unexpected increases or decreases in flow or level trends occur
• Level data are missing or incorrect but velocity data are valid
English
15
Содержание AV9000S
Страница 2: ...English 3 Fran ais 22 Espa ol 42 Italiano 63 Deutsch 83 Portugu s 103 Polski 123 T rk e 143 2...
Страница 39: ...Fran ais 39...
Страница 59: ...Espa ol 59...
Страница 62: ...45 1 14 m 1 1 1 2 48 1 21 m 1 0 2 2 62 Espa ol...
Страница 80: ...80 Italiano...
Страница 100: ...100 Deutsch...
Страница 120: ...120 Portugu s...
Страница 140: ...140 Polski...
Страница 162: ......
Страница 163: ......