background image

 

 
 

Hach AS950 - Quick Start Guide 

 
Time Based Sampling Program  
 
After Initial Setup 

 At each new sampling location 

 

1.

 

Apply power to the sampler and turn on the controller. 

2.

 

Press MENU key on keypad 

3.

 

Navigate to Programming 

4.

 

Select Programming 

5.

 

Navigate to and select Sampler Programming 

6.

 

If program has not been modified since initial setup, you can scroll past steps 7 & 8 by pressing 
the down arrow key.  

7.

 

Select Total Bottles enter “1” and select OK

 

8.

 

Select Bottle Volume enter “3 gallon”

 and select OK 

9.

 

Select Tubing enter “

length of intake tubing to the nearest foot

” and select OK

 

10.

 

If program has not been modified since initial setup, you can scroll past steps 11 & 18 by 
pressing the down arrow key.  

11.

 

Select Pacing then select Time Weighted, select NEXT 

12.

 

With Time Weighted highlighted select Edit, enter “10 minutes” select OK

 

13.

 

Select Take First. Select Edit, select “Immediately” select Back

 

14.

 

Select Sample Volume, select Fixed 

15.

 

Select Volume enter “50 ml”, select OK, select Back

 

16.

 

Select Pr

ogram Start, select Immediately on ‘Run’, select Next

 

17.

 

Select Program End, select Edit, select After Samples, press Select 

18.

 

Highlight Samples, select Edit, enter “144” select OK, select Back

 

19.

 

Select RUN/HALT key on keypad 

20.

 

Select Start Program 

21.

 

If warned about clearing data, affirm by selecting OK 

22.

 

Sampler is now Running 
 

NOTE: 

Bold font

 in step 9 indicates a variable parameter that is site dependent. Each time the sampler 

is moved to a new location you will want to enter the length of tubing between the intake strainer and 
the sampler to the nearest foot. Entering the tubing length ensures that the sample volume will be 
within tolerance. 
 