Option
Description
Delay
Selects the sampling program to start when the start trigger occurs after a selected delay.
Refer to
on page 15 for the delay options.
Control
Sets the sampling program to start (or start and stop) when the trigger occurs. Options: Start
& Stop or Start Only.
5.
If External AUX is selected, select and configure each option.
Option
Description
Delay
Selects the sampling program to start when the signal is sent by the flow meter or a third-party
device after a selected delay. Refer to
on page 15 for the delay options.
Control
Sets the sampling program to start (or start and stop) when the signal is sent by the flow meter or
third-party device. Options: Start & Stop or Start Only.
Select when the program stops
Select when the sampling program completes and sample collection stops (e.g., after a quantity of
samples are collected or after a time interval).
1.
Push
MENU
.
2.
Go to Programming>Sample Programming>Program End.
3.
Select an option.
Option
Description
None
The sampling program operates continuously.
After Samples
The sampling program stops after a set number of samples are collected.
Date and Time
The sampling program stops at a set date and time.
Time Duration
The sampling program stops after a number of hours (1 to 999 hours).
Calibration
Volume calibration
Calibrate the sample volume.
Note: Temporarily stop the sampling program before calibration. Push RUN/HALT, then select Halt Program.
Calibrate the sample volume—liquid detector enabled
Calibrate the sample volume with the liquid detector enabled to adjust sample volumes slightly.
1.
Push
MENU
.
2.
Select Hardware Setup>Liquid Detect>Status. Make sure that the liquid detector is enabled.
3.
Put the intake tubing into the sample water or tap water.
Note: Use the sample water for best accuracy.
4.
For the AWRS sampler, put a graduated cylinder under the distributor outlet or composite tube
support inside the AWRS cabinet.
5.
For the refrigerated sampler, put a graduated cylinder under the distributor outlet or composite
tube support in the refrigerated cabinet.
6.
For the portable sampler, put the tubing from the outlet side of the pump into a graduated
cylinder.
7.
Push
CALIBRATION
, then select Volume>Sample Volume.
8.
Push
Next
.
9.
Enter the volume of sample to collect.
16
English
Содержание AS950 AWRS
Страница 235: ...2 1 3 5 2 4 6 1 MENU 2 Programming Sample Programming 3 235...
Страница 300: ...1 1 5 9 2 6 10 3 7 11 4 8 1 301 3 301 3 301 317 2 pH 322 300...
Страница 301: ...3 3 AUX AUX I O AWRS 1 2 3 301...
Страница 302: ...12 24 12 24 12 24 50 1 10 99 999 0 2 1 2 3 16 0 3 0 3 303 AUX I O AUX I O AUX I O 306 1 302...
Страница 305: ...5 US900X 0 13 3 96 5 25 150 75 0 13 3 96 10 50 156 00 2 30 10 6 AV9000S 0 00 0 30 0 00 12 00 0 61 0 61 24 00 24 00 305...
Страница 308: ...d 4 a 15 15 15 15 b c OK 1 2 3 5 6 10 12 15 20 30 60 pH 307 1 2 3 a b 4 a Temp 2 2 IO IO9000 308...
Страница 309: ...5 2 6 OK IO9000 AUX IO9000 307 2 1 3 5 2 4 6 309...
Страница 310: ...1 2 3 4 1 2 3 3 3 310 310...
Страница 311: ...6 33 9 5 3 8 15 100 1 2 3 15 1 999 100 100 AUX I O AUX Flow 1 IO9000 IO 1 IO9000 307 50 1 2 3 10 10 000 1 2 311...
Страница 313: ...AUX I O 313 7 12 7 1 12 7 AUX AUX AUX I O Flow 1 2 1 2 IO 1 IO 2 IO 1 IO 2 IO9000 IO9000 1 2 3 AUX AUX I O 313...
Страница 314: ...4 308 7 313 5 AUX 7 313 1 2 3 1 999 1 2 314...
Страница 315: ...3 4 AWRS AWRS 5 6 7 8 9 10 11 12 13 14 200 1 2 AWRS 3 3 4 5 3 6 100 200 202 1 7 8 315...
Страница 316: ...9 1 OK 10 AWRS a b c 3 5 11 3 5 12 2 11 5 1 2 3 4 AWRS AWRS 5 6 7 8 9 10 11 12 1 2 3 4 5 6 316...
Страница 317: ...7 OK 8 1 2 3 AWRS 1 2 312 1 2 317...
Страница 318: ...3 4 AWRS 4 318...
Страница 319: ...1 2 3 307 1 2 3 319...
Страница 320: ...1 2 3 4 5 308 1 2 3 4 5 6 USB USB 5 USB USB USB 5 FSDATA 320...
Страница 322: ...pH 307 AUX I O AUX I O 1 2 3 317 320 320 AUX IO9000 AUX I O 0 4 20 ADC IO9000 AUX I O IO9000 322...
Страница 323: ...323...
Страница 430: ...2 1 3 5 2 4 6 1 MENU 2 Programming Sample Programming 3 430...
Страница 440: ...Program Halted Program Complete 3 Select Waiting On None Delay Schedule Setpoint 4 AWRS 4 440...
Страница 445: ...Sensor Ports AUX IO9000 AUX I O 0 4 20 mA ADC AUX I O IO9000 IO9000 Distributor Keypad Display Memory 445...
Страница 470: ......
Страница 471: ......