3.
Select an option.
Option
Description
Immediately on
'Run'
The sampling program starts when
RUN/HALT
is pushed.
On Trigger
The sampling program starts (or starts and stops) when a trigger occurs. The trigger is a
selected channel alarm or a signal from an external flow meter or third-party device that
is connected to the AUX I/O port. Refer to
Start the program on a trigger (optional)
on page 15.
After Delay
When
RUN/HALT
is pushed, the sampling program starts after the selected delay. Refer
for the delay options. For descriptions of all the options, refer to the expanded
user manual on the manufacturer's website.
On Schedule
The sampling program starts on the selected day(s) of the week and time(s). Select a
maximum of 12 schedule intervals.
Delay
—Sets the sampling program to start on the first scheduled day and time after a
selected delay. The delay is one time only. Refer to
Schedule Intervals
—Sets the number of schedule intervals (1 to 12).
Edit Schedule
—
Sets the start day and time and the stop day and time for each schedule interval.
Table 7 Delay options
Type
Description
None
The sampling program starts after no delay.
Data and Time The sampling program starts after a set date and time.
Time Duration
The sampling program starts after a set amount of time.
Flow Duration
The sampling program starts after a set flow volume occurs. Select the source of the flow
measurement.
AUX-Pulse (or AUX-mA)
—An external flow meter or third-party device connected to the AUX
I/O port.
Flow 1 (or 2)
—A sensor connected to the Sensor 1 (or 2) port.
IO 1 (or IO 2)
—Analog
Input IO 1 (or IO 2) of the optional IO9000 module.
Note: A flow measurement that is supplied through a sensor port or the IO9000 module is only available for
selection after it is selected to be recorded to the data log.
Start the program on a trigger (optional)
To do upset sampling, configure the sampling program to start on a trigger.
1.
Push
MENU
.
2.
Go to Programming>Sample Programming>Program Start>On Trigger.
3.
Select an option.
Option
Description
Setpoint
Sets the sampling program to start when one or all of the selected channel alarms occur.
External AUX
Sets the sampling program to start when the external flow meter or a third-party device that is
connected to the AUX I/O port sends a signal to the sampler. Refer to the flow meter
documentation to configure the signal.
4.
If Setpoint is selected, select and configure each option.
Option
Description
Setpoint
Selects the channel alarm(s) for the trigger. The user-configured channel alarms show. To
configure more channel alarms, select Add New Alarm and refer to
Start Trigger
Sets the sampling program to start when one or all of the channel alarms occur. Options: All
Alarms or Any Alarm.
English
15
Содержание AS950 AWRS
Страница 235: ...2 1 3 5 2 4 6 1 MENU 2 Programming Sample Programming 3 235...
Страница 300: ...1 1 5 9 2 6 10 3 7 11 4 8 1 301 3 301 3 301 317 2 pH 322 300...
Страница 301: ...3 3 AUX AUX I O AWRS 1 2 3 301...
Страница 302: ...12 24 12 24 12 24 50 1 10 99 999 0 2 1 2 3 16 0 3 0 3 303 AUX I O AUX I O AUX I O 306 1 302...
Страница 305: ...5 US900X 0 13 3 96 5 25 150 75 0 13 3 96 10 50 156 00 2 30 10 6 AV9000S 0 00 0 30 0 00 12 00 0 61 0 61 24 00 24 00 305...
Страница 308: ...d 4 a 15 15 15 15 b c OK 1 2 3 5 6 10 12 15 20 30 60 pH 307 1 2 3 a b 4 a Temp 2 2 IO IO9000 308...
Страница 309: ...5 2 6 OK IO9000 AUX IO9000 307 2 1 3 5 2 4 6 309...
Страница 310: ...1 2 3 4 1 2 3 3 3 310 310...
Страница 311: ...6 33 9 5 3 8 15 100 1 2 3 15 1 999 100 100 AUX I O AUX Flow 1 IO9000 IO 1 IO9000 307 50 1 2 3 10 10 000 1 2 311...
Страница 313: ...AUX I O 313 7 12 7 1 12 7 AUX AUX AUX I O Flow 1 2 1 2 IO 1 IO 2 IO 1 IO 2 IO9000 IO9000 1 2 3 AUX AUX I O 313...
Страница 314: ...4 308 7 313 5 AUX 7 313 1 2 3 1 999 1 2 314...
Страница 315: ...3 4 AWRS AWRS 5 6 7 8 9 10 11 12 13 14 200 1 2 AWRS 3 3 4 5 3 6 100 200 202 1 7 8 315...
Страница 316: ...9 1 OK 10 AWRS a b c 3 5 11 3 5 12 2 11 5 1 2 3 4 AWRS AWRS 5 6 7 8 9 10 11 12 1 2 3 4 5 6 316...
Страница 317: ...7 OK 8 1 2 3 AWRS 1 2 312 1 2 317...
Страница 318: ...3 4 AWRS 4 318...
Страница 319: ...1 2 3 307 1 2 3 319...
Страница 320: ...1 2 3 4 5 308 1 2 3 4 5 6 USB USB 5 USB USB USB 5 FSDATA 320...
Страница 322: ...pH 307 AUX I O AUX I O 1 2 3 317 320 320 AUX IO9000 AUX I O 0 4 20 ADC IO9000 AUX I O IO9000 322...
Страница 323: ...323...
Страница 430: ...2 1 3 5 2 4 6 1 MENU 2 Programming Sample Programming 3 430...
Страница 440: ...Program Halted Program Complete 3 Select Waiting On None Delay Schedule Setpoint 4 AWRS 4 440...
Страница 445: ...Sensor Ports AUX IO9000 AUX I O 0 4 20 mA ADC AUX I O IO9000 IO9000 Distributor Keypad Display Memory 445...
Страница 470: ......
Страница 471: ......