Option

Description

Time
Table

Collects a sample at two or more time intervals. Select this option for time-weighted sampling.

Time Intervals

—Selects the number of time intervals. 

Edit Intervals

—Configures each time

interval.

Example:

 The time intervals selected are 50, 70 and 100 minutes. The Take First setting is set

to After Interval. The Sample End setting is set to None (continuous sampling). The sampler
collects a sample after 50, 120, 220, 270 and 340 minutes and so on after the program starts.
When the Sample End setting is set to 6 samples, a sample is collected after 50, 120, 220, 270,
340 and 440 minutes after the program starts and then the sampling program is complete and
stops.

Flow
Table

Collects a sample at two or more flow intervals. Select this option for flow-weighted sampling.
An external flow meter or an optional flow sensor is necessary. Refer to the sampler
documentation to connect a flow meter or flow sensor.

Flow Source

—Selects the source of the flow measurement signal: AUX I/O port (e.g, AUX-mA),

sensor port (e.g., Flow 1) or the optional IO9000 module (e.g., IO 1). A flow measurement
supplied through a sensor port or the IO9000 module is only available for selection if the flow
measurement is recorded to the data log. Refer to 

Configure data logging

 on page 11. 

Override

—When enabled, collects a sample even if the flow is lower than estimated. Enter the maximum
time between samples. When a flow-based sample is collected, the override timer is set to zero.

Flow Intervals

—Selects the number of flow intervals. 

Edit Intervals

—Configures the flow

intervals. 

Take First

—Sets the first sample to be collected immediately or collected after the first

interval.

Example:

 The flow intervals selected are 50, 70 and 100 gallons. The Take First setting is set to

After Interval. The Sample End setting is set to None (continuous sampling). The sampler
collects a sample after 50, 120, 220, 270 and 340 gallons and so on after the program starts.
When the Sample End setting is set to 6 samples, a sample is collected after 50, 120, 220, 270,
340 and 440 gallons after the program starts and then the sampling program is complete and
stops.

Event

Collects a sample when one or more of the selected channel alarms occur. Select this option for
upset sampling.

Event Type

—Selects the type of event. Options: Setpoint. 

Event

—Selects the channel alarm(s)

for the event trigger. To configure more channel alarms, select Add New Alarm and refer to 

Select and configure the alarms

 on page 12. 

Event Trigger

—Sets a sample to be collected

when any or all of the selected alarms occur.

Select the sample volume

Select the volume of a sample. For example, set the sampling program to collect 50 mL samples.

1.

Push 

MENU

.

2.

Go to Programming>Sample Programming>Sample Volume.

3.

Select an option.

Option

Description

Fixed

Sets the sample volume to a fixed quantity, such as 500 mL. 

Volume

—Sets the volume of each

sample (10 to 10,000 mL).

Variable

Sets the sample volume to increase linearly as the flow rate increases (or decreases) between the
low flow rate and high flow rate that is entered. An external flow meter or an optional flow sensor is
necessary. Refer to the sampler documentation to connect a flow meter or flow sensor.

Note: The liquid sensor must be enabled.

Flow Source

—Selects the source of the flow measurement signal: AUX I/O port (e.g, AUX-mA),

sensor port (e.g., Flow 1) or the optional IO9000 module (e.g., IO 1). A flow measurement supplied
through a sensor port or the IO9000 module is only available for selection if the flow measurement
is recorded to the data log. Refer to 

Configure data logging

 on page 11. 

Flow Unit

—Selects the

measurement units for flow. 

Low Flow

—Enter the actual flow rate that agrees with 4 mA. 

Low

Flow Sample

—Sets the sample volume to collect when the flow is at the low flow rate. 

High Flow

—Enter the actual high flow rate that agrees with 20 mA. 

High Flow Sample

—Sets the sample

volume to collect when the flow is at the high flow rate.
For a description of fixed time interval, variable volume sampling, refer to 

CTVV sampling

on page 16.

English

   

15

Содержание AS950 AWRS

Страница 1: ...DOC346 53 80493 AS950 Controller 06 2016 Edition 8 Operations...

Страница 2: ......

Страница 3: ...stops 20 Configure cascade sampling optional 20 Configure synchronized sampling optional 21 Calibration 21 Volume calibration 21 Calibrate the sample volume liquid detector enabled 22 Adjust the sampl...

Страница 4: ...Calibrate the liquid detector 30 Appendix 30 Manning roughness coefficients 30 Table of Contents 2...

Страница 5: ...d 900MAX samplers AWRS refrigerated and portable Refer to the sampler installation and maintenance documentation for the AS950 controller specifications installation startup and maintenance informatio...

Страница 6: ...or light description Color Status Description Red Error The controller is on One or more of the measurements the sampler is configured to use e g pH and flow is not available Refer to Troubleshooting...

Страница 7: ...m dd yyyy 24h yyyy mm dd 12h yyyy mm dd 24h Display Adjusts the brightness of the display default 50 Security Enables security for password protection When enabled the user settings that can change th...

Страница 8: ...enabled in applications where liquid level changes occur Disabled Select disabled when the sample is collected from a pressurized line or when the sampler is below the sample source which can result i...

Страница 9: ...djust Level Completes a distance measurement to use for level adjust Flow Settings Device Selects the primary device Enters the device specifications Refer to Table 7 Application Settings Filter Size...

Страница 10: ...en the measured level and the actual level Options 0 61 to 0 61 m 24 00 to 24 00 in Use to correct an offset error without having to do a calibration Flow Settings Device Selects the primary device En...

Страница 11: ...a submerged area velocity sensor is connected to Sensor 2 port select Sensor 2 port Rectangular Enter the width and the height of the channel For the US900x in addition enter the source of the veloci...

Страница 12: ...e information to a Hach flow meter when a sample is collected time stamp for the sample success or fail and the bottle number When there are two sampling programs the sample information is sent when a...

Страница 13: ...urement supplied by an external flow meter that is connected directly to the AUX I O port cannot be recorded Only the recorded measurements Show on the Status screen and the Measurement Data data log...

Страница 14: ...4 Add a channel alarm as follows a Select Channel Alarms Add New Alarm Select Channel Alarm then push Next Note The number that follows a channel alarm identifies the source sensor For example Temp 2...

Страница 15: ...mple The sample distribution among the bottles When the sampling program starts When the sampling program stops Complete all the tasks that follow to configure the sampling program Select the type of...

Страница 16: ...interval 1 Push MENU 2 Go to Programming Sample Programming Pacing 3 Select an option Option Description Time Weighted Collects a sample at a set time interval such as every 15 minutes Select this opt...

Страница 17: ...the program starts When the Sample End setting is set to 6 samples a sample is collected after 50 120 220 270 340 and 440 gallons after the program starts and then the sampling program is complete and...

Страница 18: ...h the exception of AUX Pulse Samples are only collected when the flow is between the entered low flow value and high flow value The sampler calculates the sample volume to collect based on the user se...

Страница 19: ...sample into Bottle 3 and Bottle 4 Refer to item 3 in Figure 3 Time Based Sets the sampler to put the sample s that are collected during a time interval in each bottle Options 5 minutes to 24 hours def...

Страница 20: ...les are put in the correct bottles These steps are not necessary when the sampling program is temporarily stopped and then started again Change the Program Start setting to the current date Do not cha...

Страница 21: ...delay The delay is one time only Refer to Table 11 for the delay options Schedule Intervals Sets the number of schedule intervals 1 to 12 Edit Schedule Sets the start day and time and the stop day an...

Страница 22: ...the trigger occurs Options Start Stop or Start Only 5 If External AUX is selected select and configure each option Option Description Delay Selects the sampling program to start when the signal is sen...

Страница 23: ...ss and poly are necessary connect two samplers and configure them for synchronized parallel sampling The samplers collect samples at the same time Every time the first leading sampler collects a sampl...

Страница 24: ...s 13 To verify the calibration push Verify 14 To save the calibration push Finish Adjust the sample volume calibration optional For the best sample volume accuracy adjust the sample volume calibration...

Страница 25: ...et or composite tube support in the refrigerated cabinet 6 For the portable sampler put the tubing from the outlet side of the pump into a graduated cylinder 7 Push CALIBRATION then select Volume Samp...

Страница 26: ...RUN HALT 2 Select an option Option Description Start Program Starts the sampling program Note The sampling program may not start immediately Refer to Select when the sampling program starts on page 1...

Страница 27: ...tity of samples still to be collected Bottle number s of the next sample Time or counts to the next sample Time or counts to the last sample Quantity of bottles Flow information5 Current flow and the...

Страница 28: ...e cycle Sample Timeout The sample was missed because liquid was not detected within the timeout period Pump Low Volt The sample was missed because the power supply was not sufficient to operate the pu...

Страница 29: ...he alarms that occurred on a specific date and time push Options then select an option 6 To erase the alarm log push MENU and select General Settings Clear Data Select Alarm Log then push Yes Show sys...

Страница 30: ...p Manual operation Use manual operation to collect a grab sample move the distributor arm or operate the pump Note Temporarily stop the sampling program before manual operation Push RUN HALT then sele...

Страница 31: ...r has not been recorded yet The measured values shown are the last recorded value Measurement data is recorded to the data log in accordance with the selected logging interval Make sure that the measu...

Страница 32: ...t patterns Memory Shows the percent of the controller memory used Calibrate the liquid detector If the samples composition is very different than tap water calibrate the liquid detector with the guida...

Страница 33: ...lvert with bends connections and some debris 0 011 0 013 0 014 Sewer with manholes inlet straight etc 0 013 0 015 0 017 Unfinished steel form 0 012 0 013 0 014 Unfinished smooth wood form 0 012 0 014...

Страница 34: ...zed 0 011 0 013 0 015 In cement mortar 0 012 0 015 0 018 Masonry Cemented rubble 0 017 0 025 0 030 Dry rubble 0 023 0 032 0 035 Asphalt Smooth 0 013 0 013 Rough 0 016 0 016 Vegetal lining 0 030 0 500...

Страница 35: ......

Страница 36: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...

Отзывы: