Option

Description

4mA Value

Enters the flow volume that agrees with a 4 mA input at the AUX I/O port.
Configure the analog output (flow output signal) of the flow meter to have the same flow value
for 4 mA.

20mA Value

Enters the flow volume that agrees with a 20 mA input at the AUX I/O port.
Configure the analog output (flow output signal) of the flow meter to have the same flow value
for 20 mA.

7.

When Flow Reading is set to AUX-mA, select Calibrate and obey the screen prompts.
To supply 4 mA and 20 mA to the AUX I/O port:

a.

Connect the auxillary multi-purpose half cable to the AUX I/O port.

b.

Connect the orange wire (analog input) and blue wire (common) of the cable to an analog
signal source that can supply 4 mA and 20 mA.

Configure data logging

N O T I C E

 

Configure the hardware settings before data logging is configured.

Select the measurements (e.g., pH and level) that are recorded to the data log and how often the
measurements are saved to the data log (logging interval). Select a maximum of 16 measurements.
Measurements that can be recorded are supplied by the:

• Controller
• Sensors connected to the sensor ports
• Sensors connected to the analog inputs of the optional IO9000 module

Note: A flow measurement supplied by an external flow meter that is connected directly to the AUX I/O port cannot
be recorded.

Only the recorded measurements:

• Show on the Status screen and the Measurement Data (data log) screen.
• Can be used to configure setpoint alarms.
• Can be used for flow-based pacing
• Can be used to control when the sampling program starts (or starts and stops).

1.

Push 

MENU

.

2.

To configure sensors that are not yet connected to the sampler, select the sensor port
assignments. Select Programming>Datalog Programming>Change Port Assignments. Select the
sensor(s) that will be connected to the Sensor 1 port and the Sensor 2 port.

3.

Select a measurement to record as follows:

a.

Select Channel Logging.

b.

Select the source of the measurement.

c.

Select the measurement, then push 

Check

 to record the measurement.

d.

Push 

Save

.

4.

Select the logging Interval for the measurement as follows:

a.

Select Logging Intervals. The primary and secondary logging intervals show.
For example "15, 15" identifies that the primary logging interval is 15 minutes and the
secondary logging interval is 15 minutes.

b.

Select the source of the measurement.

c.

Enter the logging intervals, then push 

OK

. Options: 1, 2, 3, 5, 6, 10, 12, 15, 20, 30 or

60 minutes.

• Primary logging interval—The logging interval that is used when no alarm is configured for the

measurement. If an alarm is configured for the measurement, the primary logging interval is

English

   

11

Содержание AS950 AWRS

Страница 1: ...DOC346 53 80493 AS950 Controller 06 2016 Edition 8 Operations...

Страница 2: ......

Страница 3: ...stops 20 Configure cascade sampling optional 20 Configure synchronized sampling optional 21 Calibration 21 Volume calibration 21 Calibrate the sample volume liquid detector enabled 22 Adjust the sampl...

Страница 4: ...Calibrate the liquid detector 30 Appendix 30 Manning roughness coefficients 30 Table of Contents 2...

Страница 5: ...d 900MAX samplers AWRS refrigerated and portable Refer to the sampler installation and maintenance documentation for the AS950 controller specifications installation startup and maintenance informatio...

Страница 6: ...or light description Color Status Description Red Error The controller is on One or more of the measurements the sampler is configured to use e g pH and flow is not available Refer to Troubleshooting...

Страница 7: ...m dd yyyy 24h yyyy mm dd 12h yyyy mm dd 24h Display Adjusts the brightness of the display default 50 Security Enables security for password protection When enabled the user settings that can change th...

Страница 8: ...enabled in applications where liquid level changes occur Disabled Select disabled when the sample is collected from a pressurized line or when the sampler is below the sample source which can result i...

Страница 9: ...djust Level Completes a distance measurement to use for level adjust Flow Settings Device Selects the primary device Enters the device specifications Refer to Table 7 Application Settings Filter Size...

Страница 10: ...en the measured level and the actual level Options 0 61 to 0 61 m 24 00 to 24 00 in Use to correct an offset error without having to do a calibration Flow Settings Device Selects the primary device En...

Страница 11: ...a submerged area velocity sensor is connected to Sensor 2 port select Sensor 2 port Rectangular Enter the width and the height of the channel For the US900x in addition enter the source of the veloci...

Страница 12: ...e information to a Hach flow meter when a sample is collected time stamp for the sample success or fail and the bottle number When there are two sampling programs the sample information is sent when a...

Страница 13: ...urement supplied by an external flow meter that is connected directly to the AUX I O port cannot be recorded Only the recorded measurements Show on the Status screen and the Measurement Data data log...

Страница 14: ...4 Add a channel alarm as follows a Select Channel Alarms Add New Alarm Select Channel Alarm then push Next Note The number that follows a channel alarm identifies the source sensor For example Temp 2...

Страница 15: ...mple The sample distribution among the bottles When the sampling program starts When the sampling program stops Complete all the tasks that follow to configure the sampling program Select the type of...

Страница 16: ...interval 1 Push MENU 2 Go to Programming Sample Programming Pacing 3 Select an option Option Description Time Weighted Collects a sample at a set time interval such as every 15 minutes Select this opt...

Страница 17: ...the program starts When the Sample End setting is set to 6 samples a sample is collected after 50 120 220 270 340 and 440 gallons after the program starts and then the sampling program is complete and...

Страница 18: ...h the exception of AUX Pulse Samples are only collected when the flow is between the entered low flow value and high flow value The sampler calculates the sample volume to collect based on the user se...

Страница 19: ...sample into Bottle 3 and Bottle 4 Refer to item 3 in Figure 3 Time Based Sets the sampler to put the sample s that are collected during a time interval in each bottle Options 5 minutes to 24 hours def...

Страница 20: ...les are put in the correct bottles These steps are not necessary when the sampling program is temporarily stopped and then started again Change the Program Start setting to the current date Do not cha...

Страница 21: ...delay The delay is one time only Refer to Table 11 for the delay options Schedule Intervals Sets the number of schedule intervals 1 to 12 Edit Schedule Sets the start day and time and the stop day an...

Страница 22: ...the trigger occurs Options Start Stop or Start Only 5 If External AUX is selected select and configure each option Option Description Delay Selects the sampling program to start when the signal is sen...

Страница 23: ...ss and poly are necessary connect two samplers and configure them for synchronized parallel sampling The samplers collect samples at the same time Every time the first leading sampler collects a sampl...

Страница 24: ...s 13 To verify the calibration push Verify 14 To save the calibration push Finish Adjust the sample volume calibration optional For the best sample volume accuracy adjust the sample volume calibration...

Страница 25: ...et or composite tube support in the refrigerated cabinet 6 For the portable sampler put the tubing from the outlet side of the pump into a graduated cylinder 7 Push CALIBRATION then select Volume Samp...

Страница 26: ...RUN HALT 2 Select an option Option Description Start Program Starts the sampling program Note The sampling program may not start immediately Refer to Select when the sampling program starts on page 1...

Страница 27: ...tity of samples still to be collected Bottle number s of the next sample Time or counts to the next sample Time or counts to the last sample Quantity of bottles Flow information5 Current flow and the...

Страница 28: ...e cycle Sample Timeout The sample was missed because liquid was not detected within the timeout period Pump Low Volt The sample was missed because the power supply was not sufficient to operate the pu...

Страница 29: ...he alarms that occurred on a specific date and time push Options then select an option 6 To erase the alarm log push MENU and select General Settings Clear Data Select Alarm Log then push Yes Show sys...

Страница 30: ...p Manual operation Use manual operation to collect a grab sample move the distributor arm or operate the pump Note Temporarily stop the sampling program before manual operation Push RUN HALT then sele...

Страница 31: ...r has not been recorded yet The measured values shown are the last recorded value Measurement data is recorded to the data log in accordance with the selected logging interval Make sure that the measu...

Страница 32: ...t patterns Memory Shows the percent of the controller memory used Calibrate the liquid detector If the samples composition is very different than tap water calibrate the liquid detector with the guida...

Страница 33: ...lvert with bends connections and some debris 0 011 0 013 0 014 Sewer with manholes inlet straight etc 0 013 0 015 0 017 Unfinished steel form 0 012 0 013 0 014 Unfinished smooth wood form 0 012 0 014...

Страница 34: ...zed 0 011 0 013 0 015 In cement mortar 0 012 0 015 0 018 Masonry Cemented rubble 0 017 0 025 0 030 Dry rubble 0 023 0 032 0 035 Asphalt Smooth 0 013 0 013 Rough 0 016 0 016 Vegetal lining 0 030 0 500...

Страница 35: ......

Страница 36: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...

Отзывы: