Section 4
Page 33
MeterInstallation.fm
Mounting the Flow Meter
•
Temperature (colder temperatures decrease battery life)
•
Site hydraulics (second order effect caused by excessive turbulence)
The chart below provides typical expected battery life for each logger model:
You can access a real-time battery voltage check via InSight
®
software.
Select the 'Current Status' display. The chart below shows the estimated dc
voltage readings when approximately 30 days of battery life remain.
4.2
Mounting the Flow Meter
Important Note:
Before using the flow meter, insert the port/vent plug into the
top hole of the unit with a
7
/
16
” wrench.
DANGER
Not for use in hazardous
locations where combustible
gases may be present.
When mounting the flow meter, remember to:
•
Mount the meter so that the connectors face down.
The end of the
meter that contains the batteries should face up. If the connectors face up,
they may corrode and allow water to seep into the instrument. See
•
When not in use, cover the connectors with their protective caps to
prevent corrosion.
Use the appropriate manhole support bracket/spanner bar. See
Parts and Accessories on page 47
.
Table 4Estimated Battery Life
Model
Battery Life in Days
910
60
920
90
930
365
Table 5Battery Voltage when a 30-Day Battery Capacity Remains
Model
Battery Voltage (dc)
910
3.8
920
9.0
930
9.0
Содержание 910
Страница 1: ...Catalog Number 4975 Sigma Flow Meter Models 910 920 and 930 INSTRUMENT MANUAL 05 03 7ed...
Страница 2: ...Visit http www hach com...
Страница 4: ...Visit http www hach com...
Страница 14: ...Visit http www hach com...
Страница 16: ...Visit http www hach com...
Страница 20: ...Visit http www hach com...
Страница 26: ...Visit http www hach com...
Страница 38: ...Visit http www hach com...
Страница 42: ...Visit http www hach com...
Страница 46: ...Visit http www hach com...
Страница 48: ...Visit http www hach com...
Страница 54: ......