pH, High Purity Water,
continued
7.
Press
CAL.
The
display will show:
Standard 1?
.
8.
Rinse electrode in
deionized water and
blot dry.
9.
Place the electrode
in pH 7.0 buffer. Press
ENTER
. The display
shows:
Stabilizing...
Note: Temperature
variation causes
changes in buffer pH.
Hach sens
ion
meters
correct for these
changes during
calibration when using
buffers of pH 4, 6.86, 7,
and 10.
cal
FILL
LINE