pH, Water and Wastewater,
continued
Method Performance
Precision
In a single lab using an ASTM Substitute Wastewater Standard,
specification number D5905-96, and two electrodes with a
single sens
ion
2, a single operator obtained a standard
deviation of 0.004 pH. Each electrode was exposed to seven test
solutions with no rinsing in between and with a default
stability of 0.5 mV/min.
Interferences
Sodium error may occur in samples with pH 10 or higher. See
Section 4.2 on page 48.
Summary of Method
Water with relatively high conductivity typically has a fairly
high buffer capacity. Slight pH changes due to absorption of
carbon dioxide are usually not significant. If the sample
conductivity is not known and high accuracy is desired, follow
either the LIS or high purity methods.
The Platinum Series Combination pH Electrode responds to the
hydrogen ion concentration (activity) by developing an
electrical potential. At a constant temperature, this potential
varies linearly with the pH of the solution being measured. The
electrode has a free-diffusion junction that eliminates clogging
problems and has been proven to provide fast, accurate results.
Содержание 51910-00
Страница 2: ...2...
Страница 4: ...4...
Страница 6: ...6...
Страница 8: ...8...
Страница 16: ...16...
Страница 17: ...17 Section 2 Applications...
Страница 18: ...18...
Страница 26: ...26...
Страница 42: ...42...
Страница 46: ...46...
Страница 52: ...52...
Страница 54: ...54...
Страница 60: ...60...
Страница 64: ......