HACH LANGE TU5400 sc Скачать руководство пользователя страница 17

Startup

Set up the sensor

Use the setup menu to start an automatic vial cleaning (optional) and to select calibration,
verification, configuration and maintenance options.

1.

Push 

menu

.

2.

Select SENSOR SETUP.

3.

Select an option.

Option

Description

WIPE (optional)

Starts an automatic cleaning process to clean the vial if an automatic mechanical cleaning
unit is installed.

CALIBRATION

Shows calibration options: to start a calibration process, to set the calibration options (cal
curve, verification after calibration, cal reminder, output mode and more), to set the offset
and factor and to see the calibration log. Refer to the expanded manual on the
manufacturer's website. Refer to the calibration lid documentation for more calibration
procedure options.

VERIFICATION

Shows verification options. START—Starts the verification process. SETUP—Sets the
verification options. VERIF LOG— Shows the verification times. Refer to the calibration lid
documentation for more verification procedure options.

CONFIGURE

Shows the sensor configuration options: to change the location name, set the signal
average, measurement units, resolution, bubble reject, logger interval, button function and
more. Refer to the expanded manual on the manufacturer's website.

DIAG/TEST

Gives diagnostics and maintenance options. Refer to 

Maintenance

 on page 17.

Calibration

W A R N I N G

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.

When the instrument is used for EPA regulatory reporting, calibrations must be completed according
to EPA guidance documents and methodologies. Contact local authorities for compliance
regulations.
The sensor is factory calibrated and the laser light source is stable over time. The manufacturer
recommends that a calibration verification is done periodically to make sure that the system operates
as intended. The manufacturer recommends calibration at quarterly intervals or as local regulations
require and after any repair or comprehensive maintenance work.
Refer to the expanded manual on the manufacturer's website for more information about the
calibration settings. Calibrate the sensor with a StablCal standard(s), a formazin standards(s) or a
custom standard(s). The verification options are StablCal and glass rod.

Maintenance

W A R N I N G

 

Burn hazard. Obey safe handling protocols during contact with hot liquids.

English

   

17

-BETA DRAFT-

BETA

Содержание TU5400 sc

Страница 1: ...DOC023 52 90501 TU5400 sc TU5300 sc 09 2014 Edition 1 Basic User Manual BETA DRAFT B E T A...

Страница 2: ...BETA DRAFT B E T A...

Страница 3: ...NTU from 0 to 40 NTU 10 of reading from 40 to 700 NTU based on formazin primary standard Linearity Better than 1 for 0 to 40 NTU on formazin at 25 C 77 F Allows for accurate calibration at 20 NTU Repe...

Страница 4: ...ubing Tubing Materials Polyethylene polyamide or polyurethane tubing calibrated outside diameter in 0 001 0 004 in or 0 03 mm 0 1 mm Weight Installation environment Indoor or outdoor with cover Mounti...

Страница 5: ...death or serious injury W A R N I N G Indicates a potentially or imminently hazardous situation which if not avoided could result in death or serious injury C A U T I O N Indicates a potentially hazar...

Страница 6: ...er radiation when the instrument is defective and when the instrument lid is open This product complies with EN 61010 1 Safety Requirements for Electrical Equipment for Measurement Control and Laborat...

Страница 7: ...shock The instrument can cause interference in immediate proximity to televisions radios and computers The warranty does not cover improper use or wear FCC conformance for RFID This instrument may con...

Страница 8: ...overview D A N G E R Chemical or biological hazards If this instrument is used to monitor a treatment process and or chemical feed system for which there are regulatory limits and monitoring requireme...

Страница 9: ...F for the vial access cover1 11 Process head closed 5 Vial access cover 12 Cable channels 6 Inlet 13 Extension connector for more accessories 7 Outlet 14 Sensor cable 1 Use a screwdriver to change th...

Страница 10: ...instrument is rated for an altitude of 3100 m 10 710 ft maximum Use of this instrument at an altitude higher than 3100 m can slightly increase the potential for the insulation to breakdown which can...

Страница 11: ...tlet 3 Controller 6 Sensor Wall mount Install with a bracket D A N G E R Electrocution hazard Remove power from the instrument before this procedure is started To install the sensor on a wall with the...

Страница 12: ...12 English BETA DRAFT B E T A...

Страница 13: ...ted steps that follow Make sure to remove the thin plastic film from the mounting holes on the back of the instrument Install the maintenance rack Refer to the illustrated steps that follow to install...

Страница 14: ...sc controller Connect the sensor to the quick connect fitting of an sc controller Refer to Figure 5 The sc controller can be on or off Keep the connector cap for later use Figure 5 Quick connect fitt...

Страница 15: ...ntil the instrument is ready for operation again Yellow stable Warning for a time out e g during a calibration Condensation warning auto verification Contamination warning auto verification Yellow fla...

Страница 16: ...rd The sample line contains water under high water pressure that can burn skin if hot Qualified personnel must remove the water pressure and wear personal protective equipment during this procedure N...

Страница 17: ...n and more Refer to the expanded manual on the manufacturer s website DIAG TEST Gives diagnostics and maintenance options Refer to Maintenance on page 17 Calibration W A R N I N G Chemical exposure ha...

Страница 18: ...may increase the frequency of some tasks Table 3 Maintenance schedule Task 3 weeks 3 months 1 year Refer to Clean the vial with the wiper on page 20 settled water 1 X Refer to Clean the vial with the...

Страница 19: ...aced VIAL Shows the operational time of the vial Resets the counter when the vial is replaced in Replace the vial on page 20 CLEANING Shows the last cleaning date Resets the counter when the vial is c...

Страница 20: ...ded process in MAINTENANCE VIAL CLEANING Replace the vial N O T I C E Do not touch the glass of the vial Contamination on the glass can cause measurement errors To protect the glass from contamination...

Страница 21: ...English 21 BETA DRAFT B E T A...

Страница 22: ...fer to Plumb the sensor on page 16 to replace the tubing Clean the vial compartment Clean the vial compartment only when the compartment is contaminated Table 4 gives options on how to clean the vial...

Страница 23: ...BETA DRAFT B E T A...

Страница 24: ...w hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Compois 1222...