26
4. Operation
4.12. Access codes
The code 100 is for access to the calibration and programming of the instrument.
There are codes with other functions:
Code
Action
100
Main Menu.
401
Modification of the relay lag of the limits relays activation and deactivation.
Standard value: 2s. Programmable values: 0 to 99 s
413
Modification of the time for alarm activation when there is out of range reading.
Standard value: 60 s. Programmable values: 0 to 99 s/min/h
473
Modification of the delay time activation of the analogical output 4 to 20 mA after quitting Hold.
Standard value: 10 s. Programmable values: 0 to 99 s
497
Possibility to adjust the temperature for minimizing the effect of long cables.
si629 C_HACH LANGE_Ed 0510.indd ANG:26
si629 C_HACH LANGE_Ed 0510.indd ANG:26
20/5/10 12:06:59
20/5/10 12:06:59
Содержание si629 C
Страница 36: ...si629 C_HACH LANGE_Ed 0510 indd ALE 2 si629 C_HACH LANGE_Ed 0510 indd ALE 2 20 5 10 12 07 00 20 5 10 12 07 00 ...
Страница 66: ...si629 C_HACH LANGE_Ed 0510 indd ESP 2 si629 C_HACH LANGE_Ed 0510 indd ESP 2 20 5 10 12 07 11 20 5 10 12 07 11 ...
Страница 96: ...si629 C_HACH LANGE_Ed 0510 indd FRA 2 si629 C_HACH LANGE_Ed 0510 indd FRA 2 20 5 10 12 07 21 20 5 10 12 07 21 ...
Страница 126: ...si629 C_HACH LANGE_Ed 0510 indd ITA 2 si629 C_HACH LANGE_Ed 0510 indd ITA 2 20 5 10 12 07 32 20 5 10 12 07 32 ...
Страница 157: ...si629 C_HACH LANGE_Ed 0510 indd CONT 3 si629 C_HACH LANGE_Ed 0510 indd CONT 3 20 5 10 12 07 44 20 5 10 12 07 44 ...