English
45
f.
Use
JOB->LAB
to transfer the job file from the job status
menu to the laboratory.
(SD memory card must be inserted in the sc controller if this is
being used.)
2.
Process the job with the photometer (refer to
).
(If the SD memory card is being used, it must be connected to the
photometer via a USB adapter.)
a.
Use the
LINK2SC
button on the toolbar to open the job list.
b.
Select a job.
c.
Select the parameter to be processed.
d.
Analyze the comparison sample.
e.
Assign the result to the job.
f.
Use
SEND TO SC
to return the processed job.
(If the SD memory card is being used, it must be connected to
the photometer via a USB adapter.)
3.
LAN
1
:
When the default setting is active, the probe is automatically
corrected with the laboratory measurement data once the job has
been returned.
Note: LINK2SC>CONFIGURE>AUTO JOB
can be used to
deactivate automatic correction. The probe must then be corrected
manually as per transfer via SD memory card.
or
SD memory card:
The job must then be activated manually on the
controller. Proceed as follows:
a.
Insert the SD memory card in the sc controller.
b.
Select
LINK2SC
from the sc controller menu.
c.
Select
JOBS FROM CARD
from the LINK2SC menu. (Jobs
are loaded into the sc controller.)
d.
Select
JOB LIST
from the LINK2SC menu.
e.
Select the corresponding job from the list.
f.
Select
ACTIVATE JOB
. (The probe is corrected.)
Work sequence: NEW MEASUREMENT
4.
In order for Link2SC to be able to select the best correction type for
the application in question (refer to
section Introduction to matrix
), each further sensor correction must be performed via
"NEW MEASUREMENT" and not via "CREATE JOB".
a.
Select
LINK2SC
from the sc controller menu.
b.
Select
JOB LIST
from the LINK2SC menu.
c.
Select the corresponding job from the list.
d.
Select
NEW MEASUREMENT
.
e.
Select the
PARAMETER
for the new measurement.
f.
Take a comparison sample for the laboratory measurement.
g.
Use
JOB->LAB
to transfer the job file from the job status
menu to the laboratory.
(SD memory card must be inserted in the sc controller if this is
being used.)
5.
Process the job with the photometer (refer to 2. a. to f.)
6.
LAN
1
:
The probe is automatically corrected with the laboratory
measurement data once the job has been returned.
or
SD memory card:
The job must then be activated manually on the
controller. (refer to 3. a. to f.)
1. Only valid for sc1000 controllers.
Содержание DR 3900
Страница 10: ...Български 10 ...
Страница 18: ...Čeština 18 ...
Страница 26: ...Dansk 26 ...
Страница 34: ...Deutsch 34 ...
Страница 42: ...Ελληνικά 42 ...
Страница 50: ...English 50 ...
Страница 58: ...Español 58 ...
Страница 66: ...Suomi 66 ...
Страница 74: ...Français 74 ...
Страница 82: ...Hrvatski 82 ...
Страница 90: ...Magyar 90 ...
Страница 98: ...Italiano 98 ...
Страница 106: ...Nederlands 106 ...
Страница 114: ...Polski 114 ...
Страница 130: ...Română 130 ...
Страница 138: ...Slovenský jazyk 138 ...
Страница 146: ...Slovenščina 146 ...
Страница 154: ...Svenska 154 ...
Страница 162: ...Türkçe 162 ...
Страница 163: ...www hach com www hach lange com HACH LANGE GmbH 2011 All rights reserved Printed in Germany ...