230 V
115 V
!
!
!
Assembly/Connecting Up
11
8.
Assembly/Connecting Up
8.1
Connecting the platform to the guide rod
Lift left end of the platform (11) until the rubber connection
(10) at the bathbottom can be picked up and placed over the
pin at the underside of the platform.
8.2
Connecting to a mains socket
The unit should only be connected to a grounded mains sok-
ket. Before the unit is actually connected to the mains,
please compare the voltage indication on the name plate.
Voltage tolerances up to +/– 10% are acceptable.
8.3
Changing the mains plug (e.g. for Great Britain)
This should only be carried out by qualified spe-
cialist personnel!
The mains cable wires have the following colors:
Brown
=
Live
Blue
=
Neutral
Green/Yellow =
Earth
8.4
Fuses on the unit
All units are equipped with automatic thermally-triggered
fuses.
If the fuse has triggered
…
•
the fuse does not have to be exchanged – resetting suf-
fices;
•
a white marking is visible;
•
a certain cooling down time should be allowed (approx.
5 min) before the (dip) switch can be pressed again.
Do not use tools; do not use force. Both destroy the
fuse.
If the fuse should be triggered again after resetting,
the unit probably has a defect. In this case the unit
should be sent in for servicing.