6 Technical data
6.1 General data
Type designation:
Eyestar 900
Dimensions:
480 × 560 × 460 mm
Weight:
31.0 kg
6.2 Power
Input:
100 – 240 VAC, 50 – 60 Hz, 200 VA
6.3 Illumination modalities
6.3.1 OCT light source
Light source:
Wavelength tunable laser
Wavelength range:
1030 nm … 1080 nm
Power on patient’s eye
≤ 2.0 mW
6.3.2 Imaging system light source
Light source:
Infrared and warm white LEDs
Wavelength:
850 nm (IR)
warm white
6.3.3 Primary fixation target
Light source:
LEDs
Wavelength:
525 nm (green)
624 nm (red)
6.3.4 Fellow eye fixation target
Light source:
LEDs
Wavelength:
625 nm (red)
6.4 Measured variables
In-vivo repeatabilities reported in this chapter are based on N=46 eyes (one eye per
subject). 32 phakic (healthy and cataract) and 14 pseudophakic eyes were included.
In-vivo repeatability is calculated as the standard deviation of pairwise differences of
three repeated measurements per eye for all eyes. Aphakic and oil-filled eyes can
be measured with the device. The reported in vivo repeatability does not include
these eye types.
6.4.1 Central Corneal Thickness (CCT)
Measurement range:
300 - 800 μm
Display resolution:
1 μm
In vivo repeatability (1. SD):
± 1.5 μm
6.4.2 Anterior chamber depth (ACD)
Measurement range:
1.8 - 6.3 mm
Display resolution:
0.01 mm
In vivo repeatability (1. SD):
± 0.014 mm
6.4.3 Lens thickness (LT)
Measurement range:
0.5 - 6.5 mm
Display resolution:
0.01 mm
In vivo repeatability (1. SD):
± 0.015 mm