 INTRINSICALL SAFE REMOTE CONTROL 

OPERATING MANUAL 

90000076 R1                                                                                                                              Page 4 of 7 

1 DESCRIPTION 

The HY-OPTIMA Intrinsically Safe Remote Control Module is used to control the Explosion Proof In-
line Hydrogen Process Analyzer. 

The remote control is used to access the Configuration Menu, change settings and initiate 
Calibration.  If lost or damaged, spare remote controls can be purchased directly from H2scan.   

1.1 

INTRINSICALLY SAFE REQUIREMENTS 

Only replace the battery in a safe (non-hazardous) location. 

WARNING - Only the following batteries have been tested and approved for maintaining the 
intrinsic safety of the remote control: 

Panasonic CR1632 

Energizer CR1632 

Renata CR1632 

Using any other battery may impair the safety of the remote. 

A lanyard or anything else attached to the remote while in hazardous locations must not have 
metallic parts.  

1.2 OPERATION 

To control the 2700, hold the remote control so the LED points towards the 2700’s display 
panel. 

Pressing 

 (up arrow button) moves the cursor up. 

Pressing 

 (down arrow button) moves the cursor down. 

Pressing 

 (right arrow button) selects the next menu. 

Pressing 

 (left arrow button) selects the previous menu. 

Pressing 

  

(enter button) selects the function. 

See the 2700 Operator’s Manual for menu structure. 

1.3 BATTERY 

REPLACEMENT 

See the photos below. 

Pry open the case with a flat-tip screwdriver. 

Remove the board from the case, battery end out first. 

Replace the battery. Be sure the negative (-) side touches the board and the positive side (+) 
touches the battery holder. 

Place the case half with the buttons face down on a surface. 

Slide the board in LED end first, battery side up.  

There are small tabs in the case that fit into the notches on either side of the board.  The sides 
of the case may need to be bent out slightly so the board can snap in. 

Snap the case together. 

 

 

 

Содержание HY-OPTIMA

Страница 1: ...Parts of this document are controlled in a schedule drawing See 90000076 LTDS ...

Страница 2: ......

Страница 3: ...HY OPTIMA Intrinsically Safe Remote Control Module Operating Manual ...

Страница 4: ...INTRINSICALL SAFE REMOTE CONTROL OPERATING MANUAL 90000076 R1 Page 2 of 7 MISSION STATEMENT ...

Страница 5: ...n damaged due to shipping misuse accidents mishandling neglect or problems with electrical power sources Repair work performed during the warranty period does not prolong the warranty period past the original period System operation in incorrect or inappropriate environments Usage that is not in accordance with system guidelines or an operator s failure to follow manual instructions Limitation of ...

Страница 6: ... locations must not have metallic parts 1 2 OPERATION To control the 2700 hold the remote control so the LED points towards the 2700 s display panel Pressing up arrow button moves the cursor up Pressing down arrow button moves the cursor down Pressing right arrow button selects the next menu Pressing left arrow button selects the previous menu Pressing enter button selects the function See the 270...

Страница 7: ...INTRINSICALL SAFE REMOTE CONTROL OPERATING MANUAL 90000076 R1 Page 5 of 7 LED ...

Страница 8: ...HERES PT 0 GENERAL REQUIREMENTS IEC 60079 0 ELECTRICAL APPARATUS FOR EXPLOSIVE GAS ATMOSPHERES PT 0 GENERAL REQUIREMENTS Edition 4 Revision Date 2004 EN 60079 11 2007 ELECTRICAL APPARATUS FOR EXPLOSIVE GAS ATMOSPHERES PT 11 INTRINSIC SAFETY i IEC 60079 11 ELECTRICAL APPARATUS FOR EXPLOSIVE GAS ATMOSPHERES PT 11 INTRINSIC SAFETY I Edition 4 Revision Date 1999 EN 60079 26 2007 EXPLOSIVE ATMOSPHERES ...

Страница 9: ...NTROL OPERATING MANUAL 90000076 R1 Page 7 of 7 QUESTIONS PLEASE CONTACT US AT sales h2scan com Manual copyright 2011 H2scan Instrument software copyright 2011 H2scan Wide Range SensorTM protected under US patent number 5 279 795 ...

Отзывы: