Power
Maximum power 200 watts into 400 ohms (PK).
Maximum power 320 watts into 150 ohms (SP1).
Maximum power 325 watts into 200 ohms (SP2).
Maximum power 333 watts into 200 ohms (SP3).
Maximum power 128 watts into 70 ohms (DES).
Max Voltage
PK1/SP1
360 Volts Peak
PK2/SP2/TS1/TS2
434
Volts
Peak
PK3/SP3 480
Volts
Peak
T1
434
Volts
Peak
T2
480
Volts
Peak
DES
170
Volts
Peak
VP1
103
Volts
Peak
VP2
141
Volts
Peak
VP3
170
Volts
Peak
TP1
360
Volts
Peak
TP2
397
Volts
Peak
TP3
434
Volts
Peak
CAUTION
The following load curves apply to the fundamental power delivery capability of the
SuperPulse Generator alone. They do not imply a given power output for any given
instrument and connector cable configuration when used with the SuperPulse Generator.
Each accessory will self-impose an upper set power limit for the SuperPulse Generator.
Power Linearity 230V 50 Hz a.c
.
0
50
100
150
200
250
0
50
100
150
200
250
Measured Power
Set Power